What does one need to know about IEC 62304? In our 3-day 62304 Training course, we flow through the software development lifecycle drawing attention to requirements of the standard as well as related standards and always current FDA expectations from our experience.  The topics we plan to cover in our 2019 course are below. Topics: Regulatory...
Read More
The FDA released a Draft Guidance titled “Nonbinding Feedback After Certain FDA Inspections of Device Establishments”, on February 19, 2019.  It provides a very limited set of conditions where companies can request non-binding FDA feedback to the adequacy of planned corrective actions. The request must be submitted within 15 days of the issuance of the...
Read More
FDA issued a version of its “Refuse to Accept Policy for 510(k)s” dated 21-Feb-2019. This supersedes the prior 30-Jan-2018 version.  As FDA guidance goes, this a short period of time between updates.  It focuses on explaining the initial 15 day administrative completeness check, review initial questions, provision of a checklist for this administrative review which...
Read More
Company: Elekta, Inc. Date of Enforcement Report: 2/8/2019 Class II: PRODUCT MOSAIQ Oncology Information System and Sequencer MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed. Recall Number: Z-0821-2019 REASON There...
Read More
On Jan 14, 2019, FDA recognized ANSI AAMI SW91:2018 Classification of defects in health software.  The FDA recognition statement for this standard does not indicate any specific use in premarket submissions or relevant FDA guidances. It simply states it supports existing policies.  This standard is lengthy and technical in terms of its approach to defect...
Read More
Through blog posts and downloadable content, Alan Kusinitz, Sherman Eagles, Brian Pate, and other SoftwareCPR® experts keep you informed of new developments in FDA Software Regulation, enforcement actions, ISO standards related to medical devices, and also gain access to a wide variety of training aides, document templates, and checklists! Download the attached form to learn more about the different SoftwareCPR® subscriptions...
Read More
Company: Medtronic, Inc. Date of Enforcement Report: 1/17/2019 Class I: PRODUCT Medtronic’s Dual Chamber Implantable Pulse Generators (IPGs) are implanted cardiac pacemakers used to provide stimulation to increase heart rate in patients with a slow heart rhythm (bradycardia) or no heart rhythm. The pulse generator is the small implanted unit containing the battery and other...
Read More
Company: GE Healthcare, LLC Date of Enforcement Report: 2/1/2019 Class II PRODUCT GE Healthcare Centricity Universal Viewer Breast Imaging Centricity Universal Viewer is a device that displays medical images and data from various imaging sources, and from other healthcare information sources. Medical images and data can be displayed, communicated, stored, and processed. Typical users of...
Read More
Company: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Date of Enforcement Report: 2/1/2019 Class II PRODUCT MEDTRONIC CARELINK 2090 programmer Recall Number: Z-0795-2019 REASON There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its...
Read More
Company: GE Healthcare, LLC Date of Enforcement Report: 2/6/2019 Class II PRODUCT Revolution CT systems with the SmartStep Option. The system is intended for head, whole body, cardiac, and vascular X-ray Computed Tomography applications. Recall Number: Z-0776-2019 REASON On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation...
Read More
Company: NEUSOFT MEDICAL SYSTEMS IMP & EX Date of Enforcement Report: 2/5/2019 Class II PRODUCT NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY...
Read More
Company: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Date of Enforcement Report: 2/1/2019 Class II: PRODUCT MEDTRONIC CARELINK 2090 programmer Recall Number: Z-0795-2019 REASON There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its...
Read More
Company: GE Healthcare, LLC Date of Enforcement Report: 2/6/2019 Class II: PRODUCT Revolution CT systems with the SmartStep Option The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. Recall Number: Z-0776-2019 REASON On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation...
Read More
Company: NEUSOFT MEDICAL SYSTEMS IMP & EX Date of Enforcement Report: 2/5/2019 Class II: PRODUCT NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY...
Read More
Company: Becton, Dickinson and Company, BD Biosciences Date of Enforcement Report: 2/1/2019 Class III: PRODUCT BD FACSLyric(TM) 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878 BD FACSLyric(TM) is a high-performance flow cytometer designed to support both routine clinical analysis and research for the identification, quantification, and characterization of cells in support of cell...
Read More
Company: Radiometer America Inc Date of Enforcement Report: 2/5/2019 Class II: PRODUCT Product: ABL800 FLEX with Crea. The ABL800 with Crea is available in the following configurations: ABL817, ABL827 and ABL837. ABL800 Series Ref. 393-803, 393-804, 393-805, 393-806, 393-820, 393-821, 393-825, 393-826, 393-835, 393-836, 393-844, 393-845. Recall Number: Z-0801-2019 REASON From three incidents in Denmark...
Read More
Came across this website that has some very detailed “commandments” for software development at their company.  A surprisingly lengthy list of dos and don’ts related to coding, testing, designing, estimating, and managing the software lifecycle.  Does your company have anything written?  I often find that each company has some “lore” – some practices that characterize...
Read More
Datascope Corporation Product: Medical Device Date: 2/6/19 The United States Food and Drug Administration (FDA) conducted an inspection of your firm’s medical device operations at 1300 Macarthur Blvd., Mahwah, NJ, from July 30, 2018 through October 3, 2018. During the inspection, FDA investigators determined that your firm is a of intra-aortic balloon pumps (IABPs) and Cardiosave...
Read More
Company: Siemens Healthcare Diagnostics, Inc. Date of Enforcement Report: 1/29/2019 Class II PRODUCT Atellica IM 1300 Analyzer; In-vitro diagnostic testing of clinical specimens; Siemens Material Number (SMN): 1066001 Product Usage: The Atellica Solution is a multi-component system for in-vitro diagnostic testing of clinical specimens. The system is intended for the qualitative and quantitative analysis of...
Read More
The US FDA issued the final guidance for industry, “The Least Burdensome Provisions: Concept and Principles.”  This guidance is intended to accurately reflect Congress’ intent by describing the guiding principles and recommended approach for FDA staff and industry to facilitate consistent application of least burdensome principles.  FDA Least Burdensome Final Guidance
Read More
FDA posted a Stryker announcement today that the company is launching a voluntary field action (VFA) on specific units of the LIFEPAK 15 Monitor/Defibrillators. The company is notifying certain LIFEPAK 15 customers of an issue that may cause the device to lock up after a defibrillation shock is delivered. The lock-up condition is defined as...
Read More
Company: Varian Medical Systems, Inc. Date of Enforcement Report: 1/24/2019 Class II PRODUCT Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon,...
Read More
Company: Stryker Sustainability Solutions Date of Enforcement Report: 1/23/2019 Class II PRODUCT BW Lasso 2515 ANV eco Variable Diagnostic EP Catheter, REF D134301 Diagnostic electrophysiology (EP) catheters are specially designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation. Recall Number: Z-0766-2019 REASON Stryker’s Sustainability Solutions division (SSS) has...
Read More
Does FDA accept regulatory submissions for medical devices and SaMD that have software developed using agile methods? What about IEC 62304 compliance?  Can agile and lean approaches to software development be compliant? On February 18th and 19th, 2019, we will explore those topics and more at our 2019 “Being Agile & Compliant” public training course. ...
Read More
FDA is entering the test phase of its Digital Health Precertification (Pre-Cert) Pilot Program. General information on this program can be found at: https://www.fda.gov/MedicalDevices/DigitalHealth/DigitalHealthPreCertProgram/default.htm including FAQs at: https://www.fda.gov/MedicalDevices/DigitalHealth/DigitalHealthPreCertProgram/ucm577330.htm Milestones and next steps can be found at: https://www.fda.gov/MedicalDevices/DigitalHealth/DigitalHealthPreCertProgram/ucm584020.htm Generally, this program is for some of what FDA refers to as Software as a Medical Device (SaMD). ...
Read More
Company: Philips Medical Systems Nederlands Date of Enforcement Report: 1/8/2019 Class II PRODUCT Ingenia Elition S, Ingenia Elition X Recall Number: Z-0694-2019 REASON During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement. RECALLING FIRM/MANUFACTURER Philips Medical Systems Nederlands, Best, Netherlands on 12/3/2018. Voluntary:...
Read More
Company: Agfa-Gevaert, N.V. Date of Enforcement Report: 1/7/2019 Class II PRODUCT DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton...
Read More
Company: Medtronic Navigation, Inc. Date of Enforcement Report: 12/29/2018 Class II PRODUCT GE Magnetic Resonance System Various GE Magnetic Resonance System on-site software version. Product Usage: Magnetic resonance imaging equipment for general diagnostic use. Recall Number: Z-0693-2019 REASON Possible incorrect software version loaded. RECALLING FIRM/MANUFACTURER GE Healthcare, LLC, Waukesha, WI on 11/15/2018. Voluntary: Firm Initiated...
Read More
It’s nice to have medical device FDA regulations and guidance accessible without navigating through old websites or lugging printed copies around. Thankfully, SoftwareCPR’s free iOs and Android apps allow you to quickly view medical device regulations and guidance on your smartphone offline! As medical device consultants, we know from experience how important it is to...
Read More
(January 7, 2019)— John F. Murray, Jr, of Mount Airy, MD USA, has joined Crisis Prevention and Recovery LLC (DBA SoftwareCPR ®) as a partner. John retired from the US Food and Drug Administration in December 2018 after 32 years of federal service. For 25 years at FDA, John focused on FDA regulated software and...
Read More
Company: Hologic, Inc. Date of Enforcement Report: 12/27/2018 Class II PRODUCT I-View Contrast Enhanced Digital Mammography, ASY-08109 Recall Number: Z-0685-2019 REASON Calibration issue not possible to visualize contrast uptake in the subtracted mammography images RECALLING FIRM/MANUFACTURER Hologic, Inc., Danbury, CT on 12/6/2018. Voluntary: Firm Initiated recall is ongoing. VOLUME OF PRODUCT IN COMMERCE 337 units...
Read More
Company: Elekta, Inc. Date of Enforcement Report: 12/27/2018 Class II PRODUCT MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery Recall Number: Z-0686-2019 REASON If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup...
Read More
Company: Radiometer America Inc Date of Enforcement Report: 12/22/2018 Class II PRODUCT ABL800 analyzer, model numbers 393-800 and 393-801. Recall Number: Z-0682-2019 REASON The analyzer software may cause a mis-match of patient demographics and test results when a certain sequence of events occur. RECALLING FIRM/MANUFACTURER Radiometer America Inc, Brea, CA on 7/3/2018. Voluntary: Firm Initiated...
Read More
Company: Medtronic Inc. Date of Enforcement Report: 12/22/2018 Class II   PRODUCT Medtronic MiniMed 630G 6025805-003_C, Rx Only Medtronic MiniMed 670G 6025805-005_B, Rx Only Medtronic MiniMed 640G (International) Recall Number: Z-0683-2019   REASON Reports of occurrences in which insulin pumps with version 4.10 software have failed to make expected audio sounds during alerts, alarms or...
Read More
Company: Medtronic Navigation, Inc. Date of Enforcement Report: 12/29/2018 Class I: PRODUCT StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide – Instructions for cranial procedures using optical localization StealthStation Cranial version 3.0, 3.1 CE 0344 Rx Only 2018-02 9735489 Revision 3 Medtronic Navigation, Inc. Biopsy Needle Kit (9733068) Note: This device is...
Read More
Company: GE Healthcare, LLC Date of Enforcement Report: 12/19/2018 Class II PRODUCT Centricity Universal Viewer Zero Footprint Client, Model Number 2090258-001, Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2, SP9 and SP9.0.1 display o orientation markers can become is incorrect. Recall Number: Z-0641-2019 REASON Issue #1 affects Software Versions 6.0 SP7,...
Read More
Company: ICU Medical Inc Date of Enforcement Report: 12/17/2018 Class II PRODUCT SAPPHIRE Epidural Set Yellow-Striped Microbore, 0.2 Micron Filter, PAV, 123 Inch, List Number 16373-02 Sapphire Epidural Sets are indicated for the delivery of fluids from a container to a patient’s epidural space. Sets are intended for use with Sapphire Infusion Systems. Recall Number:...
Read More
Company: bioMerieux, Inc. Date of Enforcement Report: 12/19/2018 Class II PRODUCT VITEK 2 Systems Software Version 9.01 Update Kit. Recall Number: Z-0639-2019 REASON The VITEK 2 Systems Software Version 9.01 software may display incorrect organism identification information as well as the associated Antimicrobial Susceptibility Testing (AST) results when using the VITEK 2 FLEXPREP” feature. RECALLING...
Read More
Company: Ortho-Clinical Diagnostics Date of Enforcement Report: 12/20/18 Class II PRODUCT VITROS 5,1 FS Chemistry System, Version 3.0 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid. Recall Number: Z-0651-2019...
Read More
Company: Ortho-Clinical Diagnostics Date of Enforcement Report: 12/20/18 Class II PRODUCT VITROS 5600 Integrated System, Version 3.3.1 & below For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Product Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents. Recall Number:...
Read More
Company: Philips Electronics North America Corporation Date of Enforcement Report: 12/14/2018 Class II PRODUCT Xper Flex Cardio Physiomonitoring System Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 453564241911, 453564483321, 453564483331, 453564621791, 453564621801, 453564669081 Recall Number: Z-0614-2019 REASON There are 3 issues: Under certain conditions, real time waveforms may not be accurately plotted or displayed. The...
Read More
Company: Siemens Healthcare Diagnostics, Inc. Date of Enforcement Report: 12/14/2018 Class II PRODUCT Dimension Vista 500, model no. 10284473 an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. Recall Number: Z-0616-2019 (Vista 500, Model...
Read More
Company: Tosoh Bioscience Inc Date of Enforcement Report: 12/14/2018 Class II PRODUCT AIA-360 Analyzer is capable of performing two methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay. An antigen-antibody reaction begins by combining a patient sample, control, or calibrator with a diluent in an immunoreaction test cup from the...
Read More
Company: COVIDIEN LLC Date of Enforcement Report: 12/18/2018 Class I PRODUCT Puritan Bennett 980 (PB980) ventilator series: Puritan Bennett” 980 Pediatric Adult Ventilator Puritan Bennett” 980 Neonatal Ventilator Puritan Bennett” 980 Universal Ventilator Product Usage: The Puritan Bennett 980 Series Ventilator is designed for use on patient population sizes from Neonatal (NICU) through Adult who...
Read More
Here are some quick links to websites that list the harmonised standards for the different directives.
Read More
FDA posted a Medtronic announcement today that the company is launching a voluntary field action (VFA) for a software update on Puritan Bennett 980 Ventilators. The voluntary field corrective action is a software update to address customer feedback. The software updates the external USB Drive performance and its impact on Graphical User Interface (GUI) functionality...
Read More
Health Canada issued a draft guidance document titled, “Pre-market Requirements for Medical Device Cybersecurity.” Section 1.1 states: “Health Canada considers the inclusion of cybersecurity measures an important consideration in issuing medical device licenses. Therefore, this guidance document provides medical device manufacturers advice on the practices, responses and mitigation measures which can improve the cybersecurity of...
Read More
FDA released a new draft guidance today entitled, “Clarification of Radiation Control Regulations For Manufacturers of Diagnostic X-Ray Equipment Draft Guidance for Industry and Food and Drug Administration Staff“, dated December 17, 2018.  A few things to note related to software: On line 370, question 16, FDA addresses the question of the use of software...
Read More
FDA issued a report entitled “REPORT ON NON-DEVICE SOFTWARE FUNCTIONS: IMPACT TO HEALTH AND BEST PRACTICES – DECEMBER 2018.” This report contains findings on five software functions excluded from the device definition by the 21st Century Cures Act. The FDA found these software functions have more benefits than risks to patient safety and health. The...
Read More
FDA issued a draft guidance for prescription POC (Point-of-care) entitled “Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use.” This draft guidance document provides recommendations to industry about the types of information to include in their premarket submissions for blood glucose monitoring systems used for diabetes management in the health care prescription POC settings. This guidance...
Read More
1 14 15 16 17 18 94

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  June 24-26, 2025 (Virtual, Live)

For private, in-house courses, please contact us.

Email training@softwarecpr.com for more info.

 


 

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • TBD
See our post titled: 1st Quarter 2025 Agile Compliant Courses Scheduled

 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD