Tag

recall

Software Recall

Company: CareFusion 303, Inc. Date of Enforcement Report: 7/17/2024 Class II PRODUCT BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII Safe ES, Desktop PC, Wired Scanner, REF: 107-254-01; CII Safe ES, Desktop PC, Wireless Scanner, REF: 107-255-01; CII Safe ES,...
Read More
Company: Philips Respironics, Inc. Date of Enforcement Report: 5/23/2024 Class I PRODUCT OmniLab Advanced+ (OLA+) Ventilator Recall Number: Z-1817-2024 REASON: Philips Respironics, Inc. is updating use instructions for OmniLab Advanced+ (OLA+) Ventilator due to a failure in the Ventilator Inoperative alarm, which can cause therapy interruption or loss. Philips will correct the affected OLA+ to...
Read More
Company: Echonous Inc. Date of Enforcement Report: 7/10/2024 Class II PRODUCT Kosmos on iOS impacting iOS, Model Numbers: P007639, P007944, software is part of a general-purpose diagnostic ultrasound system. Recall Number: Z-2325-2024 REASON: Ultrasound system has a bug in affected iOS software, which after DICOM export, can cause inaccuracy in postprocessed measurements done in 3rd...
Read More
Company: CareFusion 303, Inc. Date of Enforcement Report: 7/3/2024 Class II PRODUCT BD Alaris System Manager REF: 9601, compatible with BD Alaris PC Units Recall Number: Z-2262-2024 REASON: Due to a software issue there is the potential that the PC unit may not connect to the server which could impact wireless data transmission to and...
Read More
Company: Siemens Healthcare Diagnostics Inc Date of Enforcement Report: 6/26/2024 Class II PRODUCT Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e Blood Gas System (Japan) 11416754; RAPIDPoint 500e Blood Gas System (ROW) 11416755 Recall Number: Z-2214-2024 REASON: Siemens Healthineers has...
Read More
Company: Blue Belt Technologies, Inc Date of Enforcement Report: 6/11/2024 Class II PRODUCT CORI Starter Kit, Real Intelligence CORI, CORI STARTER KIT – UNIVERSAL.Intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures Product Numbers: (1) ROB10000 CORI Starter Kit (2) ROB10024 Real Intelligence...
Read More
Company: Abbott Medical Group Date of Enforcement Report: 6/7/2024 Class I PRODUCT Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device Recall Number: Z-1921-2024 REASON: System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in...
Read More
Company: Nihon Kohden America Inc Date of Enforcement Report: 6/6/2024 Class II PRODUCT BSM-3000 Series Bedside Monitor REF BSM-3572A The Life Scope BSM-3000 Series Bedside Monitoring System is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility Recall Number: Z-2025-2024 REASON: Due to software issue...
Read More
Company: Verathon, Inc. Date of Enforcement Report: 6/4/2024 Class II PRODUCT Brand Name: GlideScope Core 15-inch Monitor Model/Catalog Number: 0570-0404 Product Description: Serialized Monitor Recall Number: Z-2021-2024 REASON: Due to software issues, there is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices. RECALLING...
Read More
Company: Stryker Orthopaedics Date of Enforcement Report: 5/28/2024 Class II PRODUCT Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako 3.0, 3.1. Software: TKA 2.0 + TKA 1.0.1 + THA 4.0.0.1 + PKA 3.0.2 Application Part Number: 700001590415 Recall Number: Z-1910-2024...
Read More
Company: Stryker Orthopaedics Date of Enforcement Report: 5/28/2024 Class II PRODUCT Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako 3.0, 3.1. Software: TKA 2.0 + TKA 1.0.1 + THA 4.0.0.1 + PKA 3.0.2 Application Part Number: 700001590415 Recall Number: Z-1910-2024...
Read More
Company: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Date of Enforcement Report: 5/24/2024 Class II PRODUCT Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100. Recall Number: Z-1895-2024 REASON: When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does...
Read More
Company: Stryker Leibinger GmbH & Co. KG Date of Enforcement Report: 5/23/2024 Class II PRODUCT 8000-021-002, Scopis ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic...
Read More
Company: Tornier S.A.S. Date of Enforcement Report: 5/22/2024 Class II PRODUCT Stryker Blueprint Software, Catalog #BPUE001 Recall Number: Z-1883-2024 REASON: The software bug allows for case planning with anatomic glenoid Perform / Perform Augmented implants and anatomic humeral Tornier Flex implant configurations that are incompatible and do not have regulatory approval. RECALLING FIRM/MANUFACTURER: Tornier S.A.S....
Read More
Company: Tandem Diabetes Care, Inc. Date of Enforcement Report: May 06, 2024 Class I PRODUCT t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology version 2.7 or later /UDI: 00850006613274 Recall Number: Z-1609-2024 REASON: During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the...
Read More
Company: Medtronic Navigation, Inc. Date of Enforcement Report: 5/9/2024 Class I PRODUCT Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. Recall Number: Z-1776-2024 REASON: Due to a software issue, there is the potential for...
Read More
Company: RAYSEARCH LABORATORIES AB Date of Enforcement Report: 5/1/2-24 Class II PRODUCT RayStation 8.0.0.61, 8.0.1.10. Radiation Therapy Treatment Planning System Recall Number: Z-1742-2024 REASON: Potential for reported SSD to be too high. RECALLING FIRM/MANUFACTURER: RAYSEARCH LABORATORIES AB VOLUME OF PRODUCT IN COMMERCE: 9 units DISTRIBUTION: Worldwide distribution – US Nationwide and the countries of Algeria,...
Read More
Company: SonarMed Inc Date of Enforcement Report: April 19, 2024 Class I PRODUCT AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring. Recall Number: Z-1426-2024 REASON: SonarMed Inc. is recalling the SonarMed Airway monitoring system because of a software issue resulting in failure to detect a partial obstruction within 2.5mm sensors...
Read More
The FDA announced today that a Federal District Court has entered a Consent Decree Against Philips Respironics following recall of certain sleep therapy machines. The consent decree includes key provisions aimed to prioritize patient relief and ensure the company’s regulatory compliance. The consent decree comes after Philips recalled certain ventilators, continuous positive airway pressure (CPAP)...
Read More
Company: Life Technologies Corporation Date of Enforcement Report: 3/29/2024 Class II PRODUCT Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM Dx...
Read More
Company: Landauer Date of Enforcement Report: 3/21/2024 Class II PRODUCT microStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110, and 201), 10016-000 (marketed in EMEA as 101), 10036-000, 10044-000, 10055-000, BC30058, and BC30059 Recall Number: Z-1349-2024 REASON: Permanent discontinuance of the microSTARii medical dosimetry system. Landauer has initiated this action to remove...
Read More
Company: Philips Respironics, Inc. Date of Enforcement Report: March 19, 2024 Class I PRODUCT Trilogy EVO, Trilogy EV300, Trilogy Evo O2, and Trilogy Evo Universal Recall Number: Z-1502-2024 REASON: Philips Respironics, Inc. has notified customers to update to the latest software version (1.05.06.00) of its Trilogy Evo continuous ventilator devices due to a possible power...
Read More
Company: Fresenius Kabi USA, LLC Date of Enforcement Report: 3/11/2024 Class II PRODUCT LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004 Recall Number: Z-1282-2024 REASON: Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable...
Read More
Company: Fresenius Kabi USA, LLC Date of Enforcement Report: March 7, 2024 Class I PRODUCT LVP Software of the Ivenix Infusion System (IIS), Version 5.8.0. Recall Number: Z-1484-2024 REASON: Fresenius Kabi USA, LLC. is recalling their Ivenix Infusion System (IIS), LVP Software, an infusion pump software. The recall will take the form of a software...
Read More
Company: Ortho-Clinical Diagnostics, Inc. Date of Enforcement Report: 3/1/2024 Class II PRODUCT VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished) Recall Number: Z-1224-2024 REASON: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause...
Read More
Company: PTW-FREIBURG Date of Enforcement Report: 2/22/2024 Class II PRODUCT Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology. Recall Number: Z-1181-2024 REASON: Software: If the user excludes voxels from the Gamma calculation that are below a dose...
Read More
Company: Baxter Healthcare Corporation Date of Enforcement Report: 2/16/2024 Class I PRODUCT Baxter Exactamix Pro 1200, REF EXM12DY Recall Number: Z-1001-2024 REASON: An error was identified in software versions 2.0.8 and 2.1.8 while using the “Use Some Overfill” feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery...
Read More
Company: Smiths Medical ASD, Inc. Date of Enforcement Report: 2/14/2024 Class I PRODUCT Smiths Medical, Medfusion Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder Numbers 4000-0100-50, 4000-0100-95, 4000-0101-50, 4000-0101-51, and 4000-0101-78. Designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. Recall Number: Z-2107-2012 REASON: Smiths...
Read More
Company: PHILIPS MEDICAL SYSTEMS Date of Enforcement Report: 2/8/2024 Class II PRODUCT Spectral CT on Rails: Software Version 5.1.0.X, model 728334 Recall Number: Z-1054-2024 REASON: A software issue that can incorrectly label all captured images that may lead to misdiagnosis and subsequent incorrect treatment. RECALLING FIRM/MANUFACTURER: PHILIPS MEDICAL SYSTEMS VOLUME OF PRODUCT IN COMMERCE: US:...
Read More
Company: Thoratec Corp. Date of Enforcement Report: 2/2/2024 Class I PRODUCT 1. HeartMate Touch Communication System, REF HMT1100 2. HeartMate Touch Communication System, Rental REF HMT1100-R 3. HeartMate Touch Communication System, HMT1150 The HeartMate Touch Application is loaded onto a tablet and will provide a user interface with state-of-the-art hardware/software platforms for supporting LVAD patients....
Read More
Company: Smiths Medical ASD Inc. Date of Enforcement Report: 2/1/2014 Class I PRODUCT Smith’s Medical Medfusion Model 3500 Syringe pump Recall Number: Z-0863-2024 REASON: Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusion Restarted with Incorrect Parameters, 3. Screen Lock, 4....
Read More
Company: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Date of Enforcement Report: 1/30/2024 Class II PRODUCT Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12) Recall Number: Z-0860-2024 REASON: Loss of connectivity between the FlexArm and the Table due to a software issue. RECALLING FIRM/MANUFACTURER: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. VOLUME...
Read More
Company: Philips Healthcare (Suzhou) Co., Ltd. Date of Enforcement Report: 1/5/2024 Class II PRODUCT Incisive CT, model 728143 & 728144 running Software Version 5.0.0. Recall Number: Z-0640-2024 REASON: Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan....
Read More
Company: Boston Scientific Corporation Date of Enforcement Report: 1/5/2024 Class II PRODUCT Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus....
Read More
Company: Covidien Date of Enforcement Report: 1/3/2023 Class II PRODUCT Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD Recall Number: Z-0631-2024 REASON: Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable. RECALLING FIRM/MANUFACTURER: Covidien VOLUME OF PRODUCT IN COMMERCE:...
Read More
Company: Abbott Date of Enforcement Report: 1/3/2024 Class I PRODUCT HeartMate Touch Communication System (version 1.0.32) Recall Number: Z-0842-2024 REASON: Abbott is recalling their HeartMate Touch System (version 1.0.32) for risk of unexpected pump stop or start. Issues may occur if the HeartMate Touch System is disconnected from a patient’s HeartMate Controller while a “pump...
Read More
Company: Abbott Laboratories Date of Enforcement Report: 12/22/2023 Class II PRODUCT Alinity ci-series System Control Module, REF: 03R70-01 Recall Number: There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests mechanism to...
Read More
Company: Abbott Laboratories Date of Enforcement Report: 12/22/2023 Class II PRODUCT Alinity ci-series System Control Module, REF: 03R70-01 Recall Number: Z-0613-2024 REASON: There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests...
Read More
Company: Siemens Healthcare Diagnostics Inc Date of Enforcement Report: 12/20/2023 Class II PRODUCT epoc NXS Host Blood Analysis System, Siemens Material Numbers 11413497 (EU), 11413475 (US), 11413506 (CA), 11413517 (MX), 11413518 (ROW), 11413498 (JP), 11413583 (CN, KR), 11413879 (IN) with software versions from 4.9.4 to the current version 4.11.11 Recall Number: Z-0601-2024 REASON: Siemens Healthcare...
Read More
Company: Philips North America Llc Date of Enforcement Report: 12/18/2023 Class II PRODUCT The device is a whole-body CT x-ray system. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. The system also includes signal analysis and display equipment, patient and equipment supports,...
Read More
Company: Ortho-Clinical Diagnostics, Inc. Date of Enforcement Report: 12/13/2023 Class II PRODUCT VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Version 3.8.0 Recall Number: Z-0543-2024 REASON: During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface...
Read More
Company: Medtronic Inc. Date of Enforcement Report: 12/09/2023 Class II PRODUCT Medtronic SynchroMed, Model A10 Recall Number: Z-0522-2024 REASON: The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm. RECALLING FIRM/MANUFACTURER: Medtronic Inc....
Read More
Company: Bebig Isotopentechnik Gmbh Date of Enforcement Report: 12/08/2023 Class II PRODUCT SagiPlan 2.2, CE0344, Rx Only , Brachyterapy Treatment Planning System Recall Number: Z-0517-2024 REASON: Due to software malfunction, numerical values may be rounded resulting in inaccurate measurements. RECALLING FIRM/MANUFACTURER: Bebig Isotopentechnik Gmbh VOLUME OF PRODUCT IN COMMERCE: 2 software licenses in US, 255...
Read More
Company: AURIS HEALTH INC Date of Enforcement Report: 12/01/2023 Class II PRODUCT Monarch Platform REF MON-000005-01, 100-240V- 50/60Hz 1440w Recall Number: Z-0448-2024 REASON: There is the potential that software issues may result in flipped image of the virtual Bronchoscope view. RECALLING FIRM/MANUFACTURER: AURIS HEALTH INC. VOLUME OF PRODUCT IN COMMERCE: 110 units In total DISTRIBUTION:...
Read More
Company: Philips North America Date of Enforcement Report: 11/15/2023 Class II PRODUCT Spectral CT on Rails, Software Version 5.1.0, Model Number 728334 Recall Number: Z-0306-2024 REASON: When preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be...
Read More
Company: Medtronic Navigation, Inc. Date of Enforcement Report: 11/09/2023 Class I PRODUCT StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation Cranial software, is intended as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. The StealthStation FlexENT System, with StealthStation ENT software, is intended as...
Read More
Company: Acclarent, Inc. Date of Enforcement Report: 10/20/2023 Class II PRODUCT TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z and TDC0005 Recall Number: Z-0127-2024 REASON: When using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation systems intended...
Read More
Company: EITAN MEDICAL LTD Date of Enforcement Report: 10/19/2023 Class I PRODUCT Sapphire Infusion Pumps, Models: Multi-Therapy (REF 15031-000-0028), Epidural (REF 15032-000-0027) and SapphirePlus (REF 15038-000-0001) Recall Number: Z-0094-2024 REASON: Infusion Pumps with affected software revision may fail to detect air in line, which may lead to air embolism. RECALLING FIRM/MANUFACTURER: EITAN MEDICAL LTD VOLUME...
Read More
Company: OLEA MEDICAL Date of Enforcement Report: 10/18/2023 Class II PRODUCT Functional MR V1.0 Recall Number: Z-0122-2024 REASON: When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting multiple volumes of interest, a functional magnetic resonance software bug may cause file contents and associated series descriptions to not match, which could lead to...
Read More
Company: Blue Belt Technologies, Inc Date of Enforcement Report: 10/05/2023 Class II PRODUCT Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299 Recall Number: Z-0041-2024 REASON: CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees...
Read More
1 2 3 62

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: TBD

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.