Recall – Patient monitor software error in low battery state

Company: Philips North America, LLC
Date of Enforcement Report: 6/12/2019
Class II

PRODUCT

IntelliVue MX40 Patient Monitor, Product number 865350, Service # USA: 453564262491 and 453564262511 – Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX40 does not automatically provide local monitoring or alarming when connected to the Information Center.

Recall Number: Z-1692-2019

IntelliVue MX40 Patient Monitor, Product number 865351, Service Numbers All other countries: 453564451791, 453564451811, 453564467721, 453564467741, 453564467761, 453564467781, 453564467801, 453564467821, 453564467841, 453564467861, 453564262531, and 453564262551 – Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX40 does not automatically provide local monitoring or alarming when connected to the Information Center.

Recall Number: Z-1693-2019

IntelliVue MX40 Patient Monitor, Product number 865352, Service Numbers All: 453564262571, 453564262591, 453564615311, 453564615331, 453564615351, and 453564615371 Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX40 does not automatically provide local monitoring or alarming when connected to the Information Center.

Recall Number: Z-1694-2019

IntelliVue MX40 Patient Monitor, Product number 857146, Service Numbers All: 453564711531 and 453564711541 – Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX40 does not automatically provide local monitoring or alarming when connected to the Information Center.

Recall Number: Z-1695-2019

REASON

The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State.

RECALLING FIRM/MANUFACTURER

Philips North America, LLC on 3/28/2019. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

DISTRIBUTION

US Nationwide and International

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020
Registration Link

Receive $300 discount with Premium-Individual subscription purchase (or $333 per person for Premium-Company subscription)!  Email training@softwarecpr.com
to receive discount

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020
For info on this course, email training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.