Company: Elekta, Inc. Date of Enforcement Report: 2/11/2025 Class II PRODUCT MOSAIQ Oncology Information System Recall Number: Z-1162-2025 REASON: Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale. RECALLING FIRM/MANUFACTURER: Elekta, Inc. VOLUME OF PRODUCT IN COMMERCE: 64 units DISTRIBUTION: US: PA, AL, FL,...Read More