Company:Siemens Medical Solutions USA, IncDate of Enforcement Report: 12/10/2014 Class lI: PRODUCT The Artis zee / zeego Angiography System is designed as a set of components that may be combined into different configurations to provide specialized angiography systems. Recall Number Z-0486-2015 REASON There is a potential issue on running Artis systems running software VC1x software...
Read More
Customed Inc.Product: medical devices, including convenience packs for surgical procedures Date: 1/9/2015 Failure to validate the defined user needs and intended uses of the (b)(4) used as part of the quality system according to an established protocol, as required by 21 CFR 820.70(i). Specifically, your firm implemented (b)(4) to track product inventory (incoming materials, finished/released...
Read More
Company: Abbott Molecular. Date of Enforcement Report 12/3/2014 Class lI: PRODUCT Abbott m2000sp is intended use as an automated system for performing sample preparation for nucleic acid testing. Recall Number Z-0463-2015 REASON Abbott Molecular identified some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This may cause...
Read More
Company: Toshiba American Medical Systems Inc. Date of Enforcement Report 12/3/2014 Class lI: PRODUCT Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the following CT systems: Aquilion ONE TSX-301C X-ray CT system with a 320-row 0.5 mm detector is provided with a 7.5-MHU large-capacity X-ray tube unit, and permits...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 12/3/2014 Class lI: PRODUCT IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030 Recall Number Z-0473-2015 REASON Software defect RECALLING FIRM/MANUFACTURER Philips Medical Systems, Inc., Cleveland, OH 11/9/2014. Voluntary:...
Read More
Company:Vision Rt Inc. Date of Enforcement Report 12/3/2014 Class lI: PRODUCT AlignRT- Intended for prescription use. The system is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated. Recall Number Z-0464-2015 REASON Potential failure of AlignRT to assert interlock....
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 12/3/2014 Class lI: PRODUCT BrightView product code: 882478 BrightView X product code: 882480 BrightView XCT product code: 882482 and 882454 Recall Number Z-0450-2015 REASON Software issues RECALLING FIRM/MANUFACTURER Philips Medical Systems, Inc., Cleveland, OH 11/4/2014. Voluntary: Firm Initiated recall is ongoing. VOLUME OF PRODUCT IN COMMERCE...
Read More
Company: Accuray Incorporated Date of Enforcement Report 11/26/2014 Class lI: PRODUCT CyberKnife Robotic Radiosurgery System with the first generation IRIS Variable Aperture Collimator. The CyberKnife is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. Recall Number Z-0218-2015...
Read More
Company: Elekta, Inc.. Date of Enforcement Report 11/26/2014 Class lI: PRODUCT MOSAIQ Product Usage: MOSAIQ is an oncology information system used to manage workflows for treatment planing and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed. Recall Number Z-0221-2015 REASON A problem can exist...
Read More
Company: Integra LifeSciences Corp. Date of Enforcement Report 11/19/2014 Class lI: PRODUCT Integra Licox Pt02 Monitor, Rx Only, Manufacturer: Integra LifeSciences (Ireland) Limited, IDA Business and Technology Park, Sragh, Tullamore, County Offaly, Ireland, Distributed by: Integra LifeSciences Corporation, 311 Enterprise Drive, Plainsboro, NJ 08536 Recall Number Z-0201-2015 REASON ntegra LifeSciences Corporation received 2 complaints that...
Read More
Company:Maquet Cardiovascular Us Sales, Llc. Date of Enforcement Report 11/19/2014 Class lI: PRODUCT Cardiohelp-I System Product Usage: A blood oxygenation and carbon dioxide removal system used to pump blood thorough the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 11/19/2014 Class lI: PRODUCT Philips Medical Systems Allura Xper FD20C Radiological Imaging with software version R8.2.O; System Code: 722028. The Allura Xper FD series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures....
Read More
Company: Respironics California Inc. (a division of Philips Healthcare) Date of Enforcement Report 11/19/2014 Class I: Date Recall Initiated: September 17, 2014 PRODUCT Esprit V1000 and V200 Ventilators, Model V1000 and V200, Installed with 3rd Generation Power Supplies, and 3rd Generation Power Supply Repair Part Kits ? See complete listing of serial numbers. Manufactured and...
Read More
Company: Beckman Coulter Inc.. Date of Enforcement Report 11/12/2014 Class lI: PRODUCT Access 2 Immunoassay Systems, Catalog No. 81600N, 386220, Instructions for Use Part No. B14253A. Product Usage: The Access 2 Immunoassay system is an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body...
Read More
Company: Iba Dosimetry Gmbh. Date of Enforcement Report 11/12/2014 Class lI: PRODUCT COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological. Recall Number Z-0168-2015 REASON Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over...
Read More
Company: Smiths Medical ASD, Inc. Date of Enforcement Report 11/12/2014 Class lI: PRODUCT CADD-Solis Medication Safety Software, Administrator CD, Version 3.1, REF 21-2194-0301-01. Product Usage: The CADD-Solis Medication Safety Software- Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD-Solis VIP Ambulatory Infusion Pump, CADD-Solis Ambulatory Infusion Pump,...
Read More
Company: Nellcor Puritan Bennett Inc. (dba Covidien LP). Date of Enforcement Report 11/6/2014 Class I: PRODUCT Covidien Puritan Bennett 980 Ventilator, Rx ONLY. Suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator for Neonatal (NICU) through Adult patient populations. Recall Number Z-0112-2015 REASON A software issue may lead...
Read More
Company: Merge Healthcare, Inc. Date of Enforcement Report 11/6/2014 Class lI: PRODUCT Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 10/29/2014 Class lI: PRODUCT Philips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705A Intended for: ” Monitoring the physiological parameters of pregnant women ” Non-invasive monitoring of fetal heart rates and movements. Recall Number Z-0105-2015 REASON Philips Intellivue...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 10/29/2014 Class lI: PRODUCT Philips IntelliVue Monitors with software revisions J.21.03, J.21.19. Model Product MP5 M8105A; MP5SC M8105AS; MP5T M8105AT; MP5 Upgrade M8105AU ; MP5SC Upgrade 866327 The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological...
Read More
Company:Mindray DS USA, Inc. dba Mindray North America.Date of Enforcement Report: 10/29/2014 Class lI: PRODUCT Mindray V21, Size: 515 mm X 335 mm X 685 mm, N.W.: 8 kg, G.W.: 11 kg, Qty:1 The V-Series Monitor is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The indications for use...
Read More
Company:Ortho-Clinical DiagnosticsDate of Enforcement Report: 10/22/2014 Class lI: PRODUCT VITROS 5600 Integrated System, System Product Code 6802413. For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents. Recall Number Z-0078-2015 REASON...
Read More
Company:Merge Healthcare, Inc.Date of Enforcement Report: 10/22/2014 Class lI: PRODUCT eFilm Workstation 4.0 and 4.0.1. eFilm Workstation is one of the components of a PACS (Picture Archiving and Communications System). Product Usage: eFilm Workstation is a software application that is used for viewing medical images. eFilm Workstation receives digital images and data from various sources...
Read More
Company:Ortho-Clinical DiagnosticsDate of Enforcement Report: 10/22/2014 Class lI: PRODUCT VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), System Product Code 6802445. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate,...
Read More
Company: Mckesson Information Solutions LLC.Date of Enforcement Report:10/15/2014 Class lI: PRODUCT McKesson Cardiology” Hemo intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment. Recall Number Z-0042-2015 REASON Software Error: The McKesson Cardiology Hemo calculation section incorrectly converts the Hemoglobin value before it is utilized in the applicable formula...
Read More
Company: GE Healthcare Date of Enforcement Report: 10/15/2014 Class lI: PRODUCT GE Healthcare Carescape Patient Data Module Recall Number Z-0060-2015 REASON GE Healthcare has recently become aware of a potential safety issue to the ECG calculations following a disconnect/reconnect cycle with the Patient Data Module, when used with the Carescape Bx50 monitors. If the PDM...
Read More
Company:Siemens Medical Solutions USA, IncDate of Enforcement Report: 10/7/2014 Class lI: PRODUCT Siemens Ysio Max system The Ysio Max is a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and...
Read More
Company: Philips Medical Systems (Cleveland) IncDate of Enforcement Report: 10/7/2014 Class lI: PRODUCT The Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and...
Read More
Company: Baxter Healthcare Corp..Date of Enforcement Report:10/7/2014 Class lI: PRODUCT SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids. Recall Number Z-2738-2014 REASON One Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014....
Read More
Company: Siemens Healthcare DiagnosticsDate of Enforcement Report: 10/7/2014 Class lI: PRODUCT ADVIA Centaur XP Immunoassay System (including refurbished units), This system is intended for professional use in a laboratory environment only. Tests performed using this system are intended for in vitro diagnostic use. As with all diagnostic tests, a definitive clinical diagnosis should not be...
Read More
Company:Siemens Medical Solutions USA, IncDate of Enforcement Report: 10/7/2014 Class lI: PRODUCT Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and...
Read More
Company: Siemens Healthcare DiagnosticsDate of Enforcement Report: 10/7/2014 Class lI: PRODUCT ADVIA Centaur Immunoassay System (including refurbished units), This system is intended for professional use in a laboratory environment only. Tests performed using this system are intended for in vitro diagnostic use. As with all diagnostic tests, a definitive clinical diagnosis should not be based...
Read More
Company:Siemens Medical Solutions USA, IncDate of Enforcement Report: 10/7/2014 Class lI: PRODUCT Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and...
Read More
Company: Celera Corporation.Date of Enforcement Report:10/1/2014 Class lII: PRODUCT ViroSeq HIV-1 Genotyping System v2.0, consisting of ViroSeq HIV-1 Genotyping System v2.0 reagent kit and Genotyping Software v3.0. Also contains ViroSeq HIV-1 Genotyping Software Manual and System v2.0 (3130) Operators Manual Recall Number B-0993-14 REASON ViroSeq HIV-1 Genotyping System v2.0, not meeting labeling claim, was distributed.....
Read More
Company:Siemens Medical Solutions USA, IncDate of Enforcement Report: 10/1/2014 Class lI: PRODUCT The Ysio Systems with software version VC10 The Ysio enables radiographic and tomographic exposures of the whole Body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Recall Number Z-2680-2014 REASON Siemens discovered that an unlikely...
Read More
Company:Siemens Medical Solutions USA, IncDate of Enforcement Report: 10/1/2014 Class lI: PRODUCT Artis zee and Artis zeego systems. x-ray, angiographic system Recall Number Z-2569-2014 REASON There is a potential problem with Artis zee and Artis zeego systems running software version VC21B and being used in conjunction with the Large Display, in that under certain circumstances,...
Read More
Company: Spacelabs Healthcare Inc.Date of Enforcement Report:10/1/2014 Class lII: PRODUCT Spacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. The Models are 91367, 91369, 91370, 91387, and 91393 are sold with Perioperative Mode (option D), which includes Start case/End case functionality, Upgrade Kit 040-1548-00. Spacelabs Healthcare patient monitors, functioning as...
Read More
Company: Terumo BCT, Inc.Date of Enforcement Report:10/1/2014 Class lII: PRODUCT TerumoBCT, Lakewood, CO 80215 USA. Catalog No: 6100 Spectra Optia Apheresis System Recall Number B-0995-14 REASON Spectra Optia Apheresis Systems with version 11 software, associated with unexpected behavior regarding rinseback, were distributed. RECALLING FIRM/MANUFACTURER Terumo BCT, Inc., Lakewood, CO on 8/27/2014 Voluntary: Firm Initiated recall...
Read More
Company: Dawning Technologies, IncDate of Enforcement Report: 9/24/2014 Class lI: PRODUCT ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package that provides connectivity between in-vitro diagnostic instruments (IVDs) and Laboratory Information Systems (LIS) aiding the flow of laboratory information, including test orders and results, specimen information, and patient...
Read More
Company: Biofire DefenseDate of Enforcement Report: 9/24/2014 Class lI: PRODUCT JBAIDS Instrument System Software version 3.5.0.72. The JBAIDS instrument is a portable thermocycler and real-time fluorimeter capable of simultaneous and rapid identification of multiple biological warfare and clinical pathogens from a variety of biological specimens. The software and instrument provides an integrated environment for performing...
Read More
Company: Siemens Healthcare Diagnostics, Inc.. Date of Enforcement Report: 9/24/2014 Class lI: PRODUCT Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. Recall Number Z-2665-2014 REASON Siemens Healthcare Diagnostics has identified a...
Read More
Company: MITSUBISHI HEAVY INDUSTRIES, LTD.Date of Enforcement Report: 9/24/2014 Class lI: PRODUCT VERO (TM) Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated. Recall Number Z-2653-2014 REASON Software Anomaly: Due...
Read More
Company: Varian Medical Systems, Inc.Date of Enforcement Report: 9/24/2014 Class lI: PRODUCT 4D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatment delivery by monitoring linear accelerator parameters and preventing the radiation therapy device from commencing irradiation when any parameter is out of conformance with the treatment plan. Recall Number...
Read More
Company: Ab SciexDate of Enforcement Report: 9/17/2014 Class lI: PRODUCT AB SCIEX API 3200MD” LC/MS/MS System with software: MultiQuantMD 3.0In vitro diagnostic to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens Part Number 5024543. Recall Number Z-2609-2014 REASON Software- AB SCIEX API 3200MD LC/MS/MS SYSTEM with MultiQuantMD 3.0 software may...
Read More
Company:Siemens Medical Solutions USA, IncDate of Enforcement Report: 9/17/2014 Class lI: PRODUCT Siemens Luminos dRF, Ysio or Uroskop Omnia systems. The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Recall Number Z-2292-2014 REASON It was discovered that during a RAD examination using Siemens Luminos...
Read More
Company: Radiometer America Inc..Date of Enforcement Report: 9/17/2014 Class lI: PRODUCT RADIANCE v. 2.4X, 2.5X and 2.6X Product Usage: RADIANCE is a data processing application for clinical use intended to monitor, collect, store, retrieve and process laboratory data. RADIANCE is also intended to control clinical analyzers providing laboratory data. Recall Number Z-2597-2014 REASON No reason...
Read More
Company: Illumina, Inc.Date of Enforcement Report: 9/17/2014 Class lI: PRODUCT Illumina Worklist Manager (IWM) (software v1.0.15), a component of Illumina MiSeqDx Platform. Product Usage: The Illumina MiSeqDx is a sequencing instrument that measures fluorescence signals of labeled nucleotides through the use of instrument specific reagents and flow cells (MiSeqDx Universal Kit 1.0), imaging hardware, and...
Read More
Company: GE Healthcare Date of Enforcement Report: 9/17/2014 Class lI: PRODUCT CARESCAPE Respiratory Modules E-sCO, E-sCOV, E-sCAiO, E-sCAiOV, E-sCAiOE, E-sCAiOVE; the Airway Gas Option N-CAiO and respective service exchange units. The affected modules could be in use with any of the following GE host devices: -CARESCAPE respiratory modules with CARESCAPE Monitors B850, B650 and B450,...
Read More
Company: Roche Diagnostics Operations, Inc.Date of Enforcement Report: 9/10/2014 Class lI: PRODUCT Roche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and process laboratory data. Recall Number Z-2563-2014 REASON It has been internally found that if cobas IT 1000 receives an ADT ADMIT, TRANSFER or MERGE event message that...
Read More
Company: Elekta Inc.Date of Enforcement Report: 9/10/2014 Class lI: PRODUCT Monaco The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two or three-dimensional radiation dose distributions inside patients for given...
Read More
1 19 20 21 22 23 59

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  June 24-26, 2025 (Virtual, Live)

For private, in-house courses, please contact us.

Email training@softwarecpr.com for more info.

 


 

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • TBD
See our post titled: 1st Quarter 2025 Agile Compliant Courses Scheduled

 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.