Cl II VITROS 5600 Integrated System

Company:Ortho-Clinical Diagnostics
Date of Enforcement Report: 10/22/2014
Class lI:

PRODUCT

VITROS 5600 Integrated System, System Product Code 6802413. For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Recall Number Z-0078-2015

REASON
Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied with VITROS Chemistry Products Calibrator Kits 1, 2, 3, 4, 6, or 9, an unexpected assay calibration may occur if assay targets are unassigned (i.e., hidden).

RECALLING FIRM/MANUFACTURER
Ortho-Clinical Diagnostics, Rochester, NY on 9/8/2014 Voluntary: Firm Initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
USA: 843; Foreign: 945

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software

ISO 14971 Risk Management (typically with heavy emphasis on software aspect)

IEC 81001-5-1 Cybersecurity Lifecycle

ISO 13485 QMS Requirements & optionally ISO 19011 for leading audits

Being Agile Yet Compliant (AAMI TIR 45)

IEC 29119-1 Software and System Testing

US Classification for Software Devices

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.