Cl II Iba Dosimetry Gmbh COMPASS

Company: Iba Dosimetry Gmbh.
Date of Enforcement Report 11/12/2014
Class lI:

PRODUCT

COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological. Recall Number Z-0168-2015

REASON
Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.

RECALLING FIRM/MANUFACTURER
Iba Dosimetry Gmbh, Schwarzenbruck, DE on 9/25/2014. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
124 units

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.