Company: Roche Diagnostics Corp/Hitachi Ltd Date of Enforcement Report 8/20/2008 Class:ll PRODUCT a) Roche/Hitachi Modular Analytics System, Modular D2400 Module GMMI clinical chemistry analyzer; Catalog number 04998618001. Intended for qualitative and quantitative in vitro determinations using a wide variety of tests; Z-0866-2008; b) Roche/Hitachi Modular Analytics System, Modular P800 Module GMMI clinical chemistry analyzer; Catalog...
Read More
Company: BD Biosciences Date of Enforcement Report 8/13/2008 Class:ll PRODUCT BD FACSDiva Software v6.0 and 6.1 with the HP XW4400 Computer System Model/Catalog number 642685 Software is used to simplify acquisition of flow cytometry data, Recall # Z-1530-2008 REASON Values might not update in statistics views after certain functions are performed in a worksheet. RECALLING...
Read More
Company: Beckman Coulter Inc Date of Enforcement Report 8/13/2008 Class:ll PRODUCT Cytomics FC 500 Flow Cytometry System with Data Innovations Instrument Manager, Part Numbers 626553 and 6605628 A system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles, Recall # Z-1529-2008 REASON Labeling provided by two integrated software...
Read More
Company: Hospira, Inc Date of Enforcement Report 8/13/2008 Class:ll PRODUCT Hospira LifeCare PCA Label Utility, Version 1.0 CD. The installation/user guide is entitled “LifeCare PCA Barcode Label Utility Program.” This software program is designed to allow entry of each drug concentration into a formulation library on a pharmacy computer for standardized printing of barcode labels...
Read More
Company: Biotronik, Inc Date of Enforcement Report 8/6/2008 Class:lll PRODUCT Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model Number: HF-T; Catalog Number 355263; Software Version 603.U; Catalog Model Number 359251; ICS 3000 units are composed of three Catalog/Model Numbers: 349527, 349530, and 355547. The model 349530 is the component which is loaded/updated with...
Read More
Company: Varian Oncology Date of Enforcement Report 8/6/2008 Class:ll PRODUCT High-Definition 120 MultiLeaf Collimator, Medical charged-particle radiation therapy system; Software Version 7.0; Product Code: H56; Accessory to Medical Linear Accelerator, Recall # Z-1459-2008 REASON Dose Calculation Error: Software anomaly may result in failure of a MLC leaf to reach planned position, potentially resulting in misadministration...
Read More
Company: GE Healthcare Integrated IT Solutions Date of Enforcement Report 7/23/2008 Class:ll PRODUCT GE Healthcare Centricity Perinatal (formerly QS) System – PFILS (Patient File Server) software; automatic patient data management providing clinical information at the bedside in Labor & Delivery, Mother-Baby and the Neonatal Intensive Care Unit, Recall # Z-2037-2008 REASON Patient information mix-up: When...
Read More
Company: Sunquest Information Systems, Inc., Date of Enforcement Report 8/6/2008 Class:ll PRODUCT Sunquest Laboratory Blood Bank and Blood Donor Modules v5.2, Distributed with Sunquest Laboratory v5.3 and v5.4. and Sunquest Laboratory Blood Bank and Blood Donor Modules v6.0.1, v6.0.2 and v6.4.0 (Beta), Distributed with Sunquest Laboratory v6.1, v6.2, v6.3 and v6.3.1 (Beta), Recall # B-1137-08...
Read More
Company: Beckman Coulter Inc. Date of Enforcement Report 7/30/2008 Class:ll PRODUCT a) Cytomics FC 500 Flow Cytometry System with CXP Software Version 2.1 and 2.2 – FC 500 with UPS, Automated differential cell counter; Part Number: 626553; Recall # Z-2016-2008; b) Cytomics FC 500 Flow Cytometry System with CXP Software Version 2.1 and 2.2 –...
Read More
Company: Hitachi Medical Corporation Date of Enforcement Report 7/30/2008 Class:lll PRODUCT EUB-5500 VISION 5500 Diagnostic Ultrasound Scanner w/Step 5 (V05-7) Software. Imaging device intended to provide the physician with physiological and clinical information, Recall # Z-1571-2008 REASON Miscalculation reading. A software error in the firm’s EUB-5500/HI VISION diagnostic scanning system causes a miscalculation of the...
Read More
Company: AGFA Corp Date of Enforcement Report 7/9/2008 Class:ll PRODUCT Impax 4.5, Impax 5.0, Impax 5.2 and Impax 5.2x Picture Archiving Systems (PACs) with integrated Voxar 3D software; and Impax 5.0 and Impax 5.2 PACs with integrated MediCAD software. The softwares are used to provide digital X-ray images for pre-surgical orthopedic planning, Z-1307-2008 REASON Misidentification:...
Read More
Company: Zoll Medical Date of Enforcement Report 7/2/2008 Class:ll PRODUCT Zoll E Series Defibrillator with 12 Lead ECG, Automated external defibrillator, Software Version 3.00 and higher, Recall # Z-1167-2008 REASON Incorrect Patient Records: Patient records that have been stored, then later transmitted or printed may not be the correct record. Records that are transmitted or...
Read More
Company: AGFA Corp Date of Enforcement Report 7/2/2008 Class:ll PRODUCT Agfa HeartLab Cardiovascular DICOMstore, Model # L9M8E. The software is used in the Agfa HeartLab Cardiovascular System, Recall # Z-1241-2008 REASON Misidentification: The electrocardiogram (ECG) data of one patient is misidentified as the data of another patient. RECALLING FIRM/MANUFACTURER Recalling Firm: AGFA Corp., Greenville, SC,...
Read More
Company: GE Healthcare Clinical Systems Date of Enforcement Report 6/18/2008 Class:ll PRODUCT GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2, Recall # Z-1155-2008 REASON Inaccurate guidelines: The BE9C is an optional, small radius Bi-planar convex probe intended for use with the LOGIQ 3 Expert system in Urology application. An attempt...
Read More
Company: Beckman Coulter Inc Date of Enforcement Report 5/28/2008 Class:ll PRODUCT a) Synchron LX20 Clinical Chemistry System, Part Numbers: 466200,473032, Recall # Z-1094-2008; b) Synchron LX20 PRO, Part Number 476100, Recall # Z-1095-2008; c) Synchron LXi 725, Part Number A08357, Recall # Z-1096-2008; d) UniCel DxC 600 Part Number: A10405, Recall # Z-1097-2008; e) UniCel...
Read More
Company: Sunquest Information Systems, Inc Date of Enforcement Report 5/21/2008 Class:ll PRODUCT Sunquest Laboratory Blood Bank and Blood Donor, version 6.0.1 and 6.0.2, Recall # B-1337-08 REASON Software, with a glitch or defect, was distributed. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ, by e-mail on March 26, 2008. Firm initiated recall is ongoing. VOLUME...
Read More
Company: Siemens Medical Solutions, Inc., Date of Enforcement Report 5/21/2008 Class:lll PRODUCT Syngo US Workplace Picture Archiving and Communication System, software version 2.0, 94043. Material number: 10035829, Recall # Z-1138-2008 REASON Inaccurate results: A software bug may cause inaccurate wall motion abnormality scoring results to be displayed. RECALLING FIRM/MANUFACTURER Siemens Medical Solutions, Inc., Mountain View,...
Read More
Company: Philips Healthcare Informatics, Inc., Date of Enforcement Report 5/14/2008 Class:lll PRODUCT iSite PACS image management system; Picture Archiving and Communications System; Radiology software package, Recall # Z-1141-2008 REASON Incorrect display of one of the patient’s images. This malfunction could cause an image to be incorrectly displayed, for example flipped left to right and anterior...
Read More
Company: Beckman Coulter, Inc Date of Enforcement Report 5/7/2008 Class:lll PRODUCT a) Cytomics FC 500 MPL Flow Cytometry System with MXP Software, Software Version 2.1, Automated differential cell counter, Part Number: 626554, Recall # Z-0958-2008; b) Cytomics FC 500 MPL Flow Cytometry System with CXP Software, Software Version 2.2, Automated differential cell counter, Part Number...
Read More
Company: Siemens Medical Solutions, USA, Inc Date of Enforcement Report 5/7/2008 Class:ll PRODUCT a) Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle radiation therapy system with software version 2.1, Recall # Z-0971-2008; b) Siemens Coherence Therapist system, Part No: 7339125, medical charged-particle radiation therapy system with software version 2.1, Recall # Z-0972-2008; c) Siemens...
Read More
Company: Siemens Medical Solutions USA, Inc., Date of Enforcement Report 5/7/2008 Class:ll PRODUCT Leonardo Workstation Picture Archiving and Communication System, Model Number: 7129534, Software Version: VB30B, Recall # Z-1038-2008 REASON Incorrect display: The orientation labels will be incorrectly displayed on the reconstructed InSpace 3-D image if the orientation was not originally HFS on the acquisition...
Read More
Company: GE Healthcare Integrated IT Solutions Date of Enforcement Report 5/7/2008 Class:ll PRODUCT GE Centricity PACS RA1000 Workstation; for diagnostic image analysis, Recall # Z-1104-2008 REASON Report missing characters: Four special characters, when entered into the exam notes, are not transferred to the preview panel or hard copy printout (greater than [>], less than [
Read More
Company: ACUSON Sequoia Diagnostic Ultrasound System Date of Enforcement Report 4/30/2008 Class:ll PRODUCT ACUSON Sequoia Diagnostic Ultrasound System, general purpose Diagnostic Ultrasound system, a) Material Number 8245875; b) Material Number 8246951; c) Material Number 8267697; Recall Z-0661-2008 REASON Measurement Error affecting all Sequoia systems using the PAL video standard configuration and have a perspective PC...
Read More
Company: BioRad BioPlex 2200 Software Date of Enforcement Report 4/23/2008 Class:ll PRODUCT BioRad BioPlex 2200 Software, for BioPlex 2200 Instrument, an automated immunodiagnostic analyzer, Recall # Z-0575-2008 REASON Incorrect values: An error was found in the BioPlex 2200 Software, Version 2.0 that results in assignment values being utilized for the BioPlex assay Calibrator Set that...
Read More
Recipient: Polymer Technology Systems, Inc.Product: CardioChek meters Date: 4/23/2008 The inspection revealed that the devices are adulterated under section 501(h) of the Act, 21 U.S.C. 351(h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformance with the Current Good Manufacturing Practice...
Read More
Company: LifeScan OneTouch Data Management Software Version 1.0 Date of Enforcement Report 4/16/2008 Class:lll PRODUCT LifeScan OneTouch Data Management Software Version 1.0, Part Number: 021-722-01(for OneTouch Select and UltraMini (Download) Blood Glucose Meters), Recall # Z-1317-2008 REASON Meter Temporarily Freezes– A software compatibility issue may cause the blood glucose meter to cease operations and freeze...
Read More
Recipient:iScreen, LLCProduct: Class II ophthalmic camera Date: 10/2/2007 This inspection revealed your ophthalmic cameras are adulterated within the meaning of Section 501(h) of the Act [21 USC 351(h)], in the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the Current Good Manufacturing...
Read More
Company: Siemens Medical Solutions Date of Enforcement Report 4/9/2008 Class:ll PRODUCT a) Mevatron M2/PRIMUS MID Medical Linear Accelerator, Material Number: 019490035, with control console software versions 6.5/7.2, 8.0, 9.0 and 9.1, Recall # Z-1394-2008; b) Mevatron KD2 Medical Linear Accelerator, Material Number: 01940753, with control console software versions 6.5/7.2, 8.0, 9.0 and 9.1, Recall #...
Read More
Company: Ventana Image Analysis System (VIAS), Date of Enforcement Report 3/26/2008 Class:ll PRODUCT Ventana Image Analysis System (VIAS) – PATHWAY anti-HER-2/neu (4B5) primary antibody with software version 2.3.0.2, Model Number: VIAS-799-10000-00, Pathology Image Analysis System, Recall # Z-0571-2008 REASON Incomplete upgrade: The software upgrade from February 1, 2007 may not have been completed. This software...
Read More
Company: BD Biosciences, Date of Enforcement Report 3/12/2008 Class:ll PRODUCT BD Biosciences MultiSET Flow Cytometry Software Versions 2.1, 2.1.1, and 2.2 with OS X operating systems, Recall # Z-0570-2008 REASON Inaccurate results: Software error results in inaccurate display result statistics. If the user adjusts the lymph gate or attractors in the Lab Report view without...
Read More
Company: Roche Diagnostics Corp Date of Enforcement Report 3/12/2008 Class:ll PRODUCT Roche COBAS 6000 Clinical Chemistry analyzer; Catalog Number: 04745914001, Recall # Z-0576-2008 REASON Incorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported. RECALLING FIRM/MANUFACTURER Recalling Firm: Roche Diagnostics Corp.,...
Read More
Company: Siemans- ACUSON Aspen Diagnostic Ultrasound System Date of Enforcement Report 3/12/2008 Class:ll PRODUCT Siemans- ACUSON Aspen Diagnostic Ultrasound System, general purpose Diagnostic Ultrasound system, Model Number: 8247887, Recall # Z-0719-2008 REASON Software error: Error affects all Aspen customer systems using the PAL video standard configuration and has a Perspective PC installed. This can lead...
Read More
Company: Sunquest Information Systems, Inc., Date of Enforcement Report 2/11/2008 Class:ll PRODUCT Sunquest Laboratory Blood Bank and Blood Donor Modules v6.0.1 and v6.0.2 Distributed with Sunquest Laboratory v6.1, v6.2 and v6.3 REASON In both the Blood Bank and Blood Donor modules, a problem has been identified in Blood Order Processing. The system does not trigger...
Read More
Recipient: Siemens Medical Solutions USA, Inc.. Product: molecular imaging diagnostic equipment device(s Date: 1/28/2008 This inspection revealed these devices are adulterated within the meaning of Section 501(h) of the Act [21 USC 351(h)], in the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity...
Read More
Company: Intuitive Surgical, Inc., Date of Enforcement Report 2/27/2008 Class:ll PRODUCT Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Model Number: IS 2000, Software Revision Numbers: A5.0 P5 and A5.1 P1, Recall # Z-0079-2008 REASON System Lock-Up: Software anomalies could cause product...
Read More
Company: Sunquest Information Systems, Inc., Date of Enforcement Report 1/23/2008 Class:ll PRODUCT Sunquest Laboratory Blood Bank and Blood Donor Modules v6.0.1 and v6.0.2 Distributed with Sunquest Laboratory v6.1, v6.2 and v6.3 REASON In the Sunquest Blood Bank and Blood Donor Modules, a problem has been identified in Blood Product Testing. If the user chooses to...
Read More
Company:Siemens Medical Solutions USA, Inc. Date of Enforcement Report 2/20/2008 Class:ll PRODUCT Siemans- ACUSON Aspen Diagnostic Ultrasound System, general purpose Diagnostic Ultrasound system, Model Number: 8247887, Recall # Z-0719-2008 REASON Software error: Error affects all Aspen customer systems using the PAL video standard configuration and has a Perspective PC installed. This can lead to a...
Read More
Company: Varian Medical Systems Inc, Date of Enforcement Report 2/20/2008 Class:ll PRODUCT Varian brand VARiS 1.4g (Medical Charged Particle Radiation Therapy System, Record and Verify System); Linear Accelerator with RTP Exchange v6.2, v6.6, & v8.0, All Models, Recall # Z-0720-2008 REASON Incorrect Software Validation. The ARC treatment plans containing a Dose Dynamic MLC for IMRT...
Read More
Company:Philips Medical Systems Date of Enforcement Report 2/20/2008 Class:ll PRODUCT Philips Medical Systems — Cardiac Viewer or Pulmonary Viewer software application version 2.0 in use with Precedence SPECT/CT System and all CT EBW workstations using software version 3.5, Recall Z-0676-2008 REASON Incorrect Measurement: The firm discovered a software defect in Cardiac and Pulmonary Viewer software...
Read More
Company: General Electric Medical Systems Information Technology Date of Enforcement Report 2/13/2008 Class:ll PRODUCT GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. (CIC Pro v4.1). (System hardware consists of a processing unit and a monitor), Recall # Z-0280-2008 REASON Two separate issues may occur with operating The CiC Pro v4.1 System...
Read More
Company: Novartis Vaccines and Diagnostic Date of Enforcement Report 2/13/2008 Class:ll PRODUCT Procleix NAT Tracker Version 3.0 Software, Stand Alone Blood, Recall # B-0662-08 REASON Software, with a glitch or defect, was distributed. RECALLING FIRM/MANUFACTURER Novartis Vaccines and Diagnostic, Emeryville, CA, by letter on August 2, 2007 with a follow up letter on August 9,...
Read More
Company: Sunquest Information Systems, Inc., Date of Enforcement Report 2/13/2008 Class:ll PRODUCT Sunquest Laboratory Blood Bank and Blood Donor, version 6.0.1 and 6.0.2, Recall # B-0663-08 REASON Software, with a glitch or defect, was distributed. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ, by e-mail on December 14. 2007. Firm initiated recall is ongoing. VOLUME...
Read More
Company: Varian Medical Systems, Inc., Date of Enforcement Report 2/6/2008 Class:ll PRODUCT Varian Real-Time Position Management System (RPM), A Respiratory Gating System; Version 1.7.3, All Models, Recall # Z-0554-2008 REASON Erroneously Resetting: RPM System version 1.7.3, software failure can occur that affects the gated radiotherapy treatment when phase-based gating is used. RPM 1.7 erroneously resets...
Read More
Company: Baxter Healthcare Renal Division Date of Enforcement Report 2/6/2008 Class:ll PRODUCT Baxter RenalSoft v.2.0 Patient Management Software Suite – HD Module; clinical data management software, Recall # Z-0527-2008; b) Baxter Renal Software Suite v.3.0 and v.3.1 – Renal Link HD Module; clinical data management software; Recall # Z-0528-2008 REASON Pre-Treatment/Post-Treatment Report errors: 1) The...
Read More
Company: Vital Images, Inc., Date of Enforcement Report 2/6/2008 Class:ll PRODUCT Vital Images ViTALConnect, Version 4.0 and 4.1, PET/CT Visualization option. (A web-based medical diagnostic tool that allows physicians to use PCs or notebook computers to gain remote access to 2D, 3D, and 4D advanced visualization. The software enables users to measure, rotate, and analyze...
Read More
Company: Cyberonics, Inc., , Date of Enforcement Report 2/6/2008 Class:lll PRODUCT Cyberonics, VNS Therapy System Model 250 Handheld programming system using Dell x5 handheld and preprogrammed with software v7.1 and higher, Recall # 0568-2008 REASON Screen Freezes– The Dell X5 Handheld PC screen will freeze caused due to incompatibility between the Microsoft 2002 OS and...
Read More
Recipient: Spinal, USAProduct:Slimplicity Anterior Cervical Plate System, the Spinal USA Cement Restrictor and the Spinal USA Vertebral Body Replacement (VBR) System Date:12/7/2007 This inspection revealed these devices are adulterated within the meaning of Section 501(h) of the Act [21 USC 351(h)], because the methods used in, or the facilities or controls used for, their manufacture,...
Read More
Company: Visx, Inc Date of Enforcement Report 1/30/2008 Class:ll PRODUCT Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478, Recall # Z-0365-2008 REASON Erroneous Treatment Calculations – Two software caused errors; in combination, in the WaveScan System Software, will result in an erroneous treatment calculation (overcorrection) in patients. (1) Installation of software...
Read More
Company: SCC Soft Computer Date of Enforcement Report 1/30/2008 Class:ll PRODUCT SoftBank II version 23.1, 23.1.1.x . , 23.1.2.x With Softscape 1.2.0.x, Recall # B-0505-08 REASON Software with a glitch or defect, was distributed. RECALLING FIRM/MANUFACTURER SCC Soft Computer, Clearwater, FL, by Task Management Notification and telephone on September 19, 2007. Firm initiated recall is...
Read More
Recipient:Omron Healthcare, Inc. Product:diagnostic and therapeutic foreign manufactured medical devices Date:1/18/2008 This inspection revealed, that these devices are adulterated within the meaning of Section 501(h) of the Act [21 U.S.C. 351(h)], in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with...
Read More
1 38 39 40 41 42 59

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  June 24-26, 2025 (Virtual, Live)

For private, in-house courses, please contact us.

Email training@softwarecpr.com for more info.

 


 

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • TBD
See our post titled: 1st Quarter 2025 Agile Compliant Courses Scheduled

 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.