Cl Il Hospira LifeCare PCA Label Utility

Company: Hospira, Inc
Date of Enforcement Report 8/13/2008
Class:ll

PRODUCT
Hospira LifeCare PCA Label Utility, Version 1.0 CD. The installation/user guide is entitled “LifeCare PCA Barcode Label Utility Program.” This software program is designed to allow entry of each drug concentration into a formulation library on a pharmacy computer for standardized printing of barcode labels for compounded drug product vials, Recall # Z-1181-2008

REASON
Incorrect dosage labels created: Customers have experienced independent changes in drug concentration (mg/mL to mcg/mL and vice versa) and expiration dates, and the addition of trailing zeros to the barcodes.

RECALLING FIRM/MANUFACTURER
Hospira, Inc., Lake Forest, IL, by letters dated January 3, 2008. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
127 CDs

DISTRIBUTION
Nationwide and Canada

Upcoming Training

62304, FDA, and Emerging Standards for Medical Device and HealthIT
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Sunnyvale, CA, USA
Dates:  February 4-6, 2020

QSS Software Validation
Planned Instructors:  Brian Pate, John F. Murray, Jr
Location: Boston, MA, USA
Dates:  June 2-4, 2020

To pre-register and get info on deep discounts or if you have questions, email training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.