Cl II Varian Real-Time Position Management Syste

Company: Varian Medical Systems, Inc.,
Date of Enforcement Report 2/6/2008
Class:ll

PRODUCT
Varian Real-Time Position Management System (RPM), A Respiratory Gating System; Version 1.7.3, All Models, Recall # Z-0554-2008

REASON
Erroneously Resetting: RPM System version 1.7.3, software failure can occur that affects the gated radiotherapy treatment when phase-based gating is used. RPM 1.7 erroneously resets the phase-based gating thresholds to system default during the treatment session, if the user accesses the Session options during the treatment session. This can lead to difference between the planned and delivered dose distribution for the treatment fraction accepted.

RECALLING FIRM/MANUFACTURER
Varian Medical Systems, Inc., Palo Alto, CA, by letter on September 26, 2007. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
59 units

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.