Donor Management Information System, Version 1.2. Recall #B-602-9. REASON Y2K defect in the computer software incorrectly calculates the end date of a temporary deferral to be 1900 instead of 2000, and removes the deferral from the database. CODE 11 systems were distributed. MANUFACTURER Information Data Management, Rosemont, Illinois. RECALLED BY Manufacturer, by telephone on January...Read More