Blood Bank Donor Management Class II 11 units

Donor Management Information System, Version 1.2. Recall #B-602-9.

REASON

Y2K defect in the computer software incorrectly calculates the end date of a temporary deferral to be 1900 instead of 2000, and removes the deferral from the database.

CODE

11 systems were distributed.

MANUFACTURER

Information Data Management, Rosemont, Illinois.

RECALLED BY

Manufacturer, by telephone on January 19, 1999. Firm-initiated recall complete.

DISTRIBUTION

New Jersey, Illinois, Ohio, South Carolina, Hawaii, Tennessee, Indiana, Hong Kong and the United Kingdom.

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.