Software Recall

CIBA Corning Acs:180 and ACS:180 Plus Analyzers, used to conduct various laboratory diagnostictests. Recall #Z-368-9. REASON Due to a software error, the ACS Folate and ACS T-Uptake assays may run without the required conditioning solution. As a result, the test results may be outside of the system requirements. CODE All serial numbers for the ACS...
Read More
2/14/96 ADAC Laboratories Pinnacle Radiation Therapy Planning Software 83 tapes recalled Due to a software problem, the wedge scatter fields will be calculated incorrectly if the wedge filter is not square. Software version 2.1F
Read More
Company: Siemens Medical Solutions USA, Inc. Date of Enforcement Report:3/26/2014 Class lI: PRODUCT syngo Dynamics Picture Archiving and Communication System (PACS). Model numbers 10091804, 10091805, 10091806, 10091807, 10091673 Recall Number Z-1216-2014 REASON Potential patient data mixup. Improper error handling could allow for two different patients to be displayed in syngo Dynamics. Images shown for one...
Read More