Software Recall

7/7/97 Solar 7000/8000 Patient Monitors with Software Revision 4A783 units. A software defect is causing incorrect waveform data.
Read More
McGaw Stratus Ambulatory Infusion System Model 550 and Alphamed EZ Pump, Model 550 withSoftware Version 2.32.00/2.42.01. Recall #Z-188-9. REASON The pump could under infuse or blood could back up into IV lines when the pumps are running at a slow flow rate with a large infusion set. CODE Serial Numbers: P61145, P71108, P71115, P71125, P71130,...
Read More
6/5/97 HP M1026a Anesthetic Gas Analyzer with O2 sensor2,284 units. The units may display innacurrate oxygen reading due to software problem.
Read More
3/13/97 Sulzer Intermedics Marathon SR & DR Pacemakers 3300 units RX 5000 Software Operating System A software condition caused the parameters of the 2 pacemakers to become reversed. That is, the software cannot distinguish Marathon DR (dual chamber) from Marathon SR (single-chamber) USER PRESET values; whichever values were most recently stored will be recalled.
Read More
1/10/97 Quinton Instrument Stress Test Monitors – 3 models ~7000units affected.A software anomaly could cause their treadmills to erroneously increase or decrease in speed and/or grade when the “Stop EXER” key is pressed at the end of a stage.
Read More
9/27/96 Radiation Oncology Computer Systems (ROCS) Treatment Planning System Software Version 5.0.x. An error occurs when using this software version for brachytherapy dose estimations for user specified calculation points for re-oriented linear sources.
Read More
CIBA Corning Acs:180 and ACS:180 Plus Analyzers, used to conduct various laboratory diagnostictests. Recall #Z-368-9. REASON Due to a software error, the ACS Folate and ACS T-Uptake assays may run without the required conditioning solution. As a result, the test results may be outside of the system requirements. CODE All serial numbers for the ACS...
Read More
2/14/96 ADAC Laboratories Pinnacle Radiation Therapy Planning Software 83 tapes recalled Due to a software problem, the wedge scatter fields will be calculated incorrectly if the wedge filter is not square. Software version 2.1F
Read More
Company: Siemens Medical Solutions USA, Inc. Date of Enforcement Report:3/26/2014 Class lI: PRODUCT syngo Dynamics Picture Archiving and Communication System (PACS). Model numbers 10091804, 10091805, 10091806, 10091807, 10091673 Recall Number Z-1216-2014 REASON Potential patient data mixup. Improper error handling could allow for two different patients to be displayed in syngo Dynamics. Images shown for one...
Read More
1 63 64 65

SoftwareCPR Training Courses

ISO13485:2016 ISO 13485 Internal Audit(or) Training Course (Live, 3-day)

IEC 62304 and other Emerging Standards Impacting Medical Device Software (Live, 3-day)

Being Agile & Yet CompliantISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.