Company: GE Healthcare, LLC 
Date of Enforcement Report: 10/23/2013
 Class ll:
PRODUCT
 
 GE, Optima CT660. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data.Recall Number  Z-0023-2014
REASON
 GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID, although all images are correctly annotated. This issue can cause additional images from a different patient and images from two or more patients to appear in one exam.
RECALLING FIRM/MANUFACTURER
 GE Healthcare, LLC,   Waukesha, WI on 3/26/2013. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
 380
DISTRIBUTION
 Nationwide and Internationally
___________________________________