Company: GE Healthcare, LLC
Date of Enforcement Report: 10/23/2013
Class ll:
PRODUCT
GE PET Discovery 610, PET Discovery 710. Intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging.
Recall Number Z-0024-2014
REASON
GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID, although all images are correctly annotated. This issue can cause additional images from a different patient and images from two or more patients to appear in one exam.
RECALLING FIRM/MANUFACTURER
GE Healthcare, LLC, Waukesha, WI on 3/26/2013. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
28
DISTRIBUTION
Nationwide and Internationally