Company: SenTec AG Date of Enforcement Report: 06/20/2023 Class II PRODUCT V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS). Recall Number: Z-1980-2023 REASON: The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode “production” and were not reset to “released” mode. RECALLING...