Recall – Patient monitor power failure leads to loss of audio

Company: Spacelabs Healthcare, Inc.
Date of Enforcement Report: 6/10/2020
Class II

PRODUCT

Xhibit Central, Model No. 96102 – Product Usage: intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.

Recall Number: Z-2261-2020

REASON

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

RECALLING FIRM/MANUFACTURER

Spacelabs Healthcare, Inc. on 8/30/2019. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

2197 units

DISTRIBUTION

US Nationwide and International

What’s New

FDA:  Exemptions, Variances, and Alternate Forms of Adverse Event Reporting for Medical Devices

Manufacturers, importers, and user facilities can request an exemption or variance from any or all of the medical device reporting requirements in 21 CFR Part 803.  Read more:  https://www.fda.gov/medical-devices/medical-device-reporting-mdr-how-report-medical-device-problems/exemptions-variances-and-alternate-forms-adverse-event-reporting-medical-devices