Recall – IVD Software anomaly

Company: Ortho-Clinical Diagnostics
Date of Enforcement Report: 12/20/18
Class II

PRODUCT

VITROS 5600 Integrated System, Version 3.3.1 & below For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Product Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.

Recall Number: Z-0650-2019

REASON

Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.

RECALLING FIRM/MANUFACTURER

Ortho-Clinical Diagnostics, Rochester, NY on 10/19/2018. Voluntary: Firm Initiated, recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
2413

DISTRIBUTION
U.S. Nationwide

CSV Training Course

Learn FDA expectations for software validation for computer systems, quality system software, manufacturing and production process software, and engineering tools. Email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921