Cl II SIEMENS Axiom Luminos dRF Max

Company:Siemens Medical Solutions USA, Inc.
Date of Enforcement Report: 6/3/2015
Class lI:

PRODUCT

SIEMENS Axiom Luminos dRF Max, a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.
Recall Number Z-1668-2015

REASON
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

RECALLING FIRM/MANUFACTURER
Siemens Medical Solutions USA, Inc, Malvern, PA on 4/13/2015 Voluntary: Firm Initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
133

DISTRIBUTION
Nationwide

___________________________________