3.0, 3.5 and 4.0 Cl II RaySearch RayStation 2.5

Company: RAYSEARCH LABORATORIES AB.
Date of Enforcement Report 6/3/2015
Class lI:

PRODUCT

RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Recall Number Z-1679-2015

REASON
RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.

RECALLING FIRM/MANUFACTURER
RAYSEARCH LABORATORIES AB, Stockholm, SE on 5/8/2015. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
539 units

DISTRIBUTION
Nationwide