Cl ll Varian Integrated Treatment Console 4DITC

Company: Varian Medical Systems
Date of Enforcement Report 2/2/2011
Class ll:

PRODUCT
4D Integrated Treatment Console (4DITC); All versions up to and including 8.8.17 Designed to assist the operator of a radiation device in providing accurate treatment set-ups for each patient by monitoring set up parameters and preventing the radiation therapy device from commencing irradiation while any parameter is out of conformance with the treatment plan. Model Number H48. Recall # Z-0932-2011
.

REASON
Multiple isocenter plans will load on the 4DITC and are available for imaging and treatment, but the 4DITC software does not recognize that mulitplex isocenter is present.

RECALLING FIRM/MANUFACTURER
Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, by letters on December 14, 2010. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
4188

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.