Cl ll Varian Eclipse Client software

Company: Varian Medical Systems.
Date of Enforcement Report 1/26/2011
Class ll:

PRODUCT
Varian Medical Systems, Eclipse Client software versions with build numbers from 7.3.10 through 8.6.15 (i.e. excluding 8.6.17); Model Number: H48; Radiation Treatment Planning System. Recall Z-0911-2011

REASON
Possible incorrect dose calculation: An anomaly in certain Eclipse versions where, under certain specific sequences of user actions, the dose distribution may fail to be re-calculated even though changes have been made which affect dose.

RECALLING FIRM/MANUFACTURER
Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA, by letters on August 6, 2010. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
7019