Cl ll Hill-Rom TotalCare Beds

Company: Hill Rom Inc.
Date of Enforcement Report 2/16/2011
Class ll:

PRODUCT
1) Hill-Rom TotalCare Bed, P1900. The TotalCare Bed System is intended to provide patient support ideally suited to be used in health care environments. The TotalCare Bed System may be used in a variety of settings including but not limited to, acute care, including critical care, step down/progressive care medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU) and sections of the emergency department (ED). Recall # Z-1220-2011;

2) Hill-Rom TotalCare Bariatric Bed, P1840. The TotalCare Bed System is intended to provide patient support ideally suited to be used in health care environments. The TotalCare Bed System may be used in a variety of settings including but not limited to, acute care, including critical care, step down/progressive care medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU) and sections of the emergency department (ED). Recall # Z-1221-2011.

REASON
The firm received one report that a patient was weighed on the bed at 53 Kg when the patient actually weighed 73 Kg. The patient expired the next day and it is inconclusive as to whether the scale was a contributing factor. The firm investigated and found that the software for the scale system did not clear the last patient weight from the history of the scale system.

RECALLING FIRM/MANUFACTURER
Hill-Rom, Inc., Batesville, IN, by letter dated January 19, 2011. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
945 units

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.