Day

December 10, 2003
Company: Misys Healthcare SystemsDate of Enforcement Report: 12/10/03 Class: I PRODUCT Calculator/Data Processing Module for Clinical Use. Recall # Z-0007-04. REASON Software defect. Clinical Laboratory results failing quality assurance were filed directly to a patient’s chart without review. CODE Versions 5.2, 5.23, and 5.3. RECALLING FIRM/MANUFACTURER Misys Healthcare Systems, Tucson, AZ, by facsimile on August...
Read More
Company: Misys Healthcare SystemsDate of Enforcement Report: 12/10/03 Class: II PRODUCT Misys Laboratory System. Recall # Z-0151-04. REASON Software defect causes orders to be lost under certain conditions of use. A typographical error at the Accept/Modfy/Reject prompt will cause the order not to be processed by the system. CODE Versions 5.2, 5.23, 5.3 up to...
Read More

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.