Company: Misys Healthcare SystemsDate of Enforcement Report: 12/10/03 Class: I PRODUCT Calculator/Data Processing Module for Clinical Use. Recall # Z-0007-04. REASON Software defect. Clinical Laboratory results failing quality assurance were filed directly to a patient’s chart without review. CODE Versions 5.2, 5.23, and 5.3. RECALLING FIRM/MANUFACTURER Misys Healthcare Systems, Tucson, AZ, by facsimile on August...
Read More
Company: Misys Healthcare SystemsDate of Enforcement Report: 12/10/03 Class: II PRODUCT Misys Laboratory System. Recall # Z-0151-04. REASON Software defect causes orders to be lost under certain conditions of use. A typographical error at the Accept/Modfy/Reject prompt will cause the order not to be processed by the system. CODE Versions 5.2, 5.23, 5.3 up to...