Day

September 22, 1999
IMED Gemini PC-4 Volumetric Infusion Pump and Controller withVersion 1.85 Software: a) PC-4 Infusion Pump, Model 1340, 110 volts; b) PC-4 Infusion Pump, Model 1341, 220 volts. Recall #Z-174/175-0. REASON Due to a software modification, the devices will stop infusing and alarm if the processors receive conflicting or confusing messages. CODE All pumps that carry...