6/22/99 Bayer Corporation 3767 units. Class III Technicon SETpoint Chemistry Calibrator. During review of the value assignment procedure it was learned that the operator manually entered intercept corrected SETpoint values into an Excel spreadsheet rather than letting the spreadsheet calculate the corrections from the true SETpoint. The values that were entered were incorrect and, therefore,...