Bayer Corporation 3767 units.(IVD) Class III

6/22/99 Bayer Corporation 3767 units. Class III Technicon SETpoint Chemistry Calibrator. During review of the value assignment procedure it was learned that the operator manually entered intercept corrected SETpoint values into an Excel spreadsheet rather than letting the spreadsheet calculate the corrections from the true SETpoint. The values that were entered were incorrect and, therefore, the resulting value assignments were incorrect. Z-1072-9

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.