Day

March 2, 1999
All IMMULITE 2000 Automated Immunoassay Analyzers with SoftwareVersion 1.1, a clinical automated immunoassay analyzer intended for in-vitro diagnostic use for detection of analytates in human specimens. Recall #Z-909-9. REASON If the initial counts of the reaction tubes are exactly 10,000 cps, this leads to erroneously low calculated results. CODE All serial numbers in commercial distribution...
Read More
Chronimed 3/2/99 Glucose test strips and meters 5. Failure to establish and maintain plans that describe or reference the design and development activities and define responsibility for implementation as required by 21 CFR 820.30(b). For example, there is no plan for the software validation for Version 104. 6. Failure to ensure that the design input...
Read More