August 19, 1998
Sabratek Corp. 8/19/98 Infusion pumps 1. Failure to identify the actions needed to correct and prevent recurrence of non-conforming product. For example, it was determined that the _____ had a software problem which prevented the infusion rate from exceeding the bolus rate while operating in the “PCA” mode. While the software has been revised, the...