McGHAN Medical Corporation 2/16/96 mammary prostheses The validation of the computerized dip timer system for software version [purged word] and [purged word] fails to assure that the previous files are overwritten when the maximum number of files [purged word] is reached. This is demonstrated by the fact that there was no verification of the file...