Day

February 16, 1996
McGHAN Medical Corporation 2/16/96 mammary prostheses The validation of the computerized dip timer system for software version [purged word] and [purged word] fails to assure that the previous files are overwritten when the maximum number of files [purged word] is reached. This is demonstrated by the fact that there was no verification of the file...
Read More

Upcoming Public Courses

No public courses are planned at this time.

If you would like to discuss a private, onsite course, please fill out the form below.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.