Tag

recall

Software Recall

6/25/98 Medtronic 9790C Programmers for Kappa 400 Pacemakekers Model 9952A software: Class II 133 units. Bench testing identified a software anomaly that could result in the delivery of an unitended pacing mode or pacing interval. Recall #=Z-715-8
Read More
6/24/98 Beckman/Coulter Access Immunochemistry Set 81500 Software versions 3.22 and 3.23: Class II 1296 units. Sofware anomaly in these versions may allow test results to be posted to a different patient sample number. Recall #=Z-786/787-8.
Read More
6/17/98 Lifescan Surestep Blood Glucose Monitoring System (Meter of home use) Changed to Class I 727,004 units The meters may give an Er 1 (Error 1)message if a patient’s blood glucose is 500 mg/dL or greater.
Read More
Resuscitaire Radiant Warmer, Infant Radiant Warmer: a) Resuscitaire Radiant Warmer, Model RW81; b) Resuscitaire Radiant Warmer, Model RW82; c) Resuscitaire Radiant Warmer, Model RW82VHA; d) Resuscitaire Radiant Warmer, Model WMRW82; e) Resuscitaire Radiant Warmer, Model WBR 81; f) Resuscitaire Radiant Warmer, Model WBR82. Recall #Z-292/297-9. REASON The device fails to control heater power due to...
Read More
CIBA Corning ACS 180 Plus Analyzers, used for numerous assays. Recall #Z-359-9. REASON A software error for Troponin I analysis caused erroneous results. CODE All ACS 180 Plus Analyzers with test definition AL software version. MANUFACTURER Chiron Diagnostics Corporation, Oberlin, Ohio. RECALLED BY Manufacturer, by letter/technical bulletin on June 9, 1998. Firm-initiated field correction complete....
Read More
Venturi Ventilator, provides a software that continuously adaptsto the patient’s changing breathing demands and assures ventilation at the lowest pressure. Recall #Z-803-9. REASON Software error may cause ventilator to deliver more than set tidal volume. CODE Venturi Ventilator with Software Revision C. Serial #970222. MANUFACTURER Cardiopulmonary Corporation, Milford, Connecticut. RECALLED BY Manufacturer, by letter on...
Read More
Information Management System for Mammography Practices. Recall #Z-187-9. REASON The software was manufactured with methods that resulted in defective information being provided to the physician, and therefore, the patients with positive mammograms may not follow-up with a biopsy exam. CODE All copies of Mammoworks Version 2.6. MANUFACTURER Lumisys, Inc., (formerly Compurad), Tucson, Arizona. RECALLED BY...
Read More
ACS:Centaur Automated Chemiluminescence System, used to conduct various laboratory diagnostictests. Recall #Z-363-9. REASON Software errors for V1.0 which could lead to erroneous test results. CODE All lot numbers for the ACS Centaur. MANUFACTURER Chiron Diagnostics Corporation, Oberlin, Ohio. RECALLED BY Manufacturer, by fax dated March 25, 1998. Firm-initiated field correction complete. DISTRIBUTION Germany and Italy....
Read More
3/17/98 Recon Rx neurological biofeedback monitoring1762 units Rx unit were not functioning correctly. Obsolete version 3.01 software had been installed in the units.
Read More
Venturi Ventilator, used to provide a software that continouslyadapts to the patient’s changing breathing demands and assures ventilation at the lowest pressure. The display monitor provides pertient patient parameters and waveforms. Recall #Z-802-9. REASON Display monitor may go blank due to faulty inverter. CODE Serial numbers: 970192, 970190, 970154, 970164, 970169, 970172, 970138, 970161, 970163,...
Read More
CIBA Corning ACS:180 SE Analyzer, used to conduct various laboratory diagnostic tests. Recall#Z-369-9. REASON Software error causes 16 different assays to not calculate SI units correctly. CODE All lot numbers for ACS 180 SE Analyzers. MANUFACTURER Chiron Diagnostics Corporation, Oberlin, Ohio. RECALLED BY Manufacturer, by fax on November 25, 1997. Firm-initiated field correction complete. DISTRIBUTION...
Read More
9/30/97 EZ-Flow 480 Ambulatory Infusion Pump 1058 devices. Defective software may cause over infusion when using the PCA mode. 1/9/98 confirmed a patient death due to a software logic error related to use in the patient controlled analgesia (PCA) mode caused the lethal dose of morphine sulfate. Two non-fatal incidents in 1996 also occurred.
Read More
7/7/97 Solar 7000/8000 Patient Monitors with Software Revision 4A783 units. A software defect is causing incorrect waveform data.
Read More
McGaw Stratus Ambulatory Infusion System Model 550 and Alphamed EZ Pump, Model 550 withSoftware Version 2.32.00/2.42.01. Recall #Z-188-9. REASON The pump could under infuse or blood could back up into IV lines when the pumps are running at a slow flow rate with a large infusion set. CODE Serial Numbers: P61145, P71108, P71115, P71125, P71130,...
Read More
6/5/97 HP M1026a Anesthetic Gas Analyzer with O2 sensor2,284 units. The units may display innacurrate oxygen reading due to software problem.
Read More
3/13/97 Sulzer Intermedics Marathon SR & DR Pacemakers 3300 units RX 5000 Software Operating System A software condition caused the parameters of the 2 pacemakers to become reversed. That is, the software cannot distinguish Marathon DR (dual chamber) from Marathon SR (single-chamber) USER PRESET values; whichever values were most recently stored will be recalled.
Read More
1/10/97 Quinton Instrument Stress Test Monitors – 3 models ~7000units affected.A software anomaly could cause their treadmills to erroneously increase or decrease in speed and/or grade when the “Stop EXER” key is pressed at the end of a stage.
Read More
9/27/96 Radiation Oncology Computer Systems (ROCS) Treatment Planning System Software Version 5.0.x. An error occurs when using this software version for brachytherapy dose estimations for user specified calculation points for re-oriented linear sources.
Read More
CIBA Corning Acs:180 and ACS:180 Plus Analyzers, used to conduct various laboratory diagnostictests. Recall #Z-368-9. REASON Due to a software error, the ACS Folate and ACS T-Uptake assays may run without the required conditioning solution. As a result, the test results may be outside of the system requirements. CODE All serial numbers for the ACS...
Read More
2/14/96 ADAC Laboratories Pinnacle Radiation Therapy Planning Software 83 tapes recalled Due to a software problem, the wedge scatter fields will be calculated incorrectly if the wedge filter is not square. Software version 2.1F
Read More
Company: Siemens Medical Solutions USA, Inc. Date of Enforcement Report:3/26/2014 Class lI: PRODUCT syngo Dynamics Picture Archiving and Communication System (PACS). Model numbers 10091804, 10091805, 10091806, 10091807, 10091673 Recall Number Z-1216-2014 REASON Potential patient data mixup. Improper error handling could allow for two different patients to be displayed in syngo Dynamics. Images shown for one...
Read More
1 60 61 62

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: March 7 & 28, 2024

Virtual via Zoom

Registration Link:

Register Now

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.