Tag

recall

Software Recall

REASON The software validation of version 4.01R1 a software anomaly was found that affected the proper function of a print option. PRODUCT Vitalcom ST Segment Arrhythmis Detection Software Option, an Optional ECG software feature used in conjunction with the Critikon Observer Central Station on which is installed. There are 12 models or catalog numbers of...
Read More
100 copies. Pato Brachytherapy Treatment Planning System Version 13.2 and higher. Software Implementation Error. Z-039-0/
Read More
PRODUCT Model 250 NeuroCybernetic Prostheses (NCP) Programming Software Versions a) 4/1; b) 4.4, implanted autonomic nerve stimulator for he treatment of epilepsy. REASON The label insert sheets were printed with the incorrect lot number. CODE Lot numbers a) 23952C; b) 26023C, 27137C, 27735C. Recall #Z-570/571-0. MANFACTURER Cyberonics, Inc., Inc., Houston, Texas. RECALLED BY Manufacturer, by...
Read More
IMED Gemini PC-4 Volumetric Infusion Pump and Controller withVersion 1.85 Software: a) PC-4 Infusion Pump, Model 1340, 110 volts; b) PC-4 Infusion Pump, Model 1341, 220 volts. Recall #Z-174/175-0. REASON Due to a software modification, the devices will stop infusing and alarm if the processors receive conflicting or confusing messages. CODE All pumps that carry...
Read More
223 units. 740 ventilator system and ventilator 760 system. Any Ventilator will revision J. software. Software can cause under or over delivery of oxygen. Z-170/171-0
Read More
Nucletron Plato External Beam Planning Radiation Therapy SoftwareV2.1.2 and MLC/Shape Software Module V2.3 Recall #Z-038-0. REASON Coordinates for radiation beam used in therapy are mislabeled in software. CODE Plato RTS software version V2.1.2 used with software module MLC/Shape version V2.3. MANUFACTURER Nucletron BV, The Netherlands. RECALLED BY Nucletron Corporation, Columbia, Maryland, by letter and customer...
Read More
11 systems. Acqplan Radiation Therapy Planning System. A software option for computed tomography scanners. An incorrect calculation of ISO/N dosage may result when changing parameters. Z-1236-9
Read More
7-July-99 Olympus. Class III field correction. EPROM Version 3.00 Software for the Olympus EVIS CV-140 Video System Center. The device has a malfunction due to incorrect coding of the software which can cause the software to overwrite the patient ID number with any comments that are entered by the user.. Z-1227-9IVD
Read More
30-June-99 C30-June-99 Kavo Dental 7 units. Class III. AC powered dental drilling device. Software error may cause unit to shutdown prior to reaching preset torque limit. Z-1215-9
Read More
28-June-99 Quinton Instruments 11 review stations and 7 Sequence NIS stations. Class II.Under certain specific circumstances,sequences of a single plane study, may overlap sequences of bi-plane study when viewing both on the Angiocomm Single Sequence Review Station or the Single Sequence Network Interface Station (NIS). Z-1116/1119-9
Read More
6/22/99 Bayer Corporation 3767 units. Class III Technicon SETpoint Chemistry Calibrator. During review of the value assignment procedure it was learned that the operator manually entered intercept corrected SETpoint values into an Excel spreadsheet rather than letting the spreadsheet calculate the corrections from the true SETpoint. The values that were entered were incorrect and, therefore,...
Read More
6/14/99 Cardiopulmonary Corporation 59 units. Class II Venturi Ventilator with hard drive. User interface software will not load when unit is first powered on. Z-1039-9
Read More
Defense Blood Standard System (DBSS), Blood Bank Software. Recall #B-1094-9. REASON Blood bank computer software does not identify all potential duplicate donor records. CODE Software Versions 3.00 and 3.01. MANUFACTURER Electronic Data Systems (EDS), Herndon, Virginia. RECALLED BY U.S. Department of Defense (DBSS)Composite Health Care System 2, Falls Church, Virginia, by E-mail on June 9...
Read More
Boehringer Mannheim/Hitachi 917 Disk System, a clinical chemistryanalyzer system. Recall #Z-1040-9. REASON The barcode reader can read the wrong sample barcode, mismatch the test results, and report them under the wrong sample. CODE All versions of the disk system software with an updated barcode reader. This includes the system with Serial No. 9715-05 and all...
Read More
5/21/99 VIA Medical Corporation 59 units. Class II VIA Low Volume Mode (LVM) Monitor. The devices has a software problem that can cause the monitor, under unusual circumstances, to display and print a chemistry value from a previous sampling instead of the blood chemistry value from the current sampling. Z-1041-9.
Read More
Abbott Alcyon Analyzers with Alcyon System Software Version 1.6.Recall #Z-267-0. REASON The device may give incorrect tests results or reference ranges when printing the “Complete Report (header)”, “Incomplete Report (DRAFT)”, and the “Control Results printout.” CODE List #4E73-04. MANUFACTURER Abbott Laboratories, Inc., Irving, Texas. RECALLED BY Manufacturer, by telephone on May 18 and 19, 1999,...
Read More
5/18/99 Cardiopulmonary Corporation 1 unit. Class II Venturi Ventilator Software revision C. Software error may cause ventilator to deliver more than set tidal volume. Z-803-9
Read More
4/30/99 Bayer Diagnostics 1000 units. Class II ADVIA 120 Hematology System When running the ADVIA 120 system in the manual mode using the Host Query feature and the timeout is set to 13 seconds, it is possible for the system to report incorrect patient results. The supplemental labeling instructions section of the product labeling instructs...
Read More
4/16/99 Class II Colleague 3 Volumetric Infusion Pump (Triple Channel). Baxter Healthcare 6,992 units. A software communication error occurs predominately when all three channels are in use, causing an alarm codition that will stop the function of all of the channels in use. The pump will alarm appropriately, alerting the user of the failure state....
Read More
All Architect I System Processing Modules with Software Versions1.0 and 1.01. Recall #Z-910-9. REASON The device may report incorrect results for diluted samples; and/or when used with the Laboratory Information System or Host Information Systems, the system may incorrectly associate test results with patient ID or incorrectly associate control results with patient records. CODE List...
Read More
REASON Defective software could result in failure to give warning. CODE Version V1.4 and all previous versions of the software. MANUFACTURER Marquette Medical Systems, Inc., Milwaukee, Wisconsin. RECALLED BY Manufacturer, by telephone and letter sent on April 8, 1999. Firm-initiated field correction ongoing. DISTRIBUTION Nationwide, Canada, France, Australia, Italy, Sweden. QUANTITY 368 units were distributed.
Read More
Donor Management Information System, Version 1.2. Recall #B-602-9. REASON Y2K defect in the computer software incorrectly calculates the end date of a temporary deferral to be 1900 instead of 2000, and removes the deferral from the database. CODE 11 systems were distributed. MANUFACTURER Information Data Management, Rosemont, Illinois. RECALLED BY Manufacturer, by telephone on January...
Read More
4/6/99 Class II Phillips Medical 1086 units. Integris family of x-ray controls and generators (fluoroscopy,, urology, cardiology). The defect occurs when the system is driven to maximum EER and the source to image receptor distance (SID) is moved toa shorter distance while continuing to make exposures. In this manner of operation, the output may exceed...
Read More
3/29/99 Digisonics Class II OBLink software releases prior to 9.0 39 units. Y2K sofvtware error may produce an incorrect estimated delivery date (EDD). Z-1056-9.
Read More
Eagle 3000 Floorloader Steam Sterilizer. Recall #Z-936-9. REASON There is an error in the control system software, which causes the chamber temperatures to exceed the upper limit. CODE Various serial numbers. MANUFACTURER Steris Corporation, Erie Pennsylvania. RECALLED BY Steris Corporation, Mentor, Ohio, by Customer Notifications on March 10, 1999. Firm-initiated recall ongoing. DISTRIBUTION Nationwide, Canada,...
Read More
Philips Medical Sonodiagnost 800, intended for endovaginal biopsy for various gynecologic and obstetric procedures including follicular retrieval, endometrial biopsy, and fluid aspiration of any type in a potentially pregnant patient. Recall #Z-772-9. REASON Software used with Transducer EV7014 incorrectly indicates track of the biopsy needle. CODE All SD800’s with Software versions 2.1. and 2.1.1 used...
Read More
3/5/99 Class II Hewlett Packard 145 units. Software used with transducer EV7014 incorrectly indicates track of biospy needle. Z-772-9
Read More
All IMMULITE 2000 Automated Immunoassay Analyzers with SoftwareVersion 1.1, a clinical automated immunoassay analyzer intended for in-vitro diagnostic use for detection of analytates in human specimens. Recall #Z-909-9. REASON If the initial counts of the reaction tubes are exactly 10,000 cps, this leads to erroneously low calculated results. CODE All serial numbers in commercial distribution...
Read More
Abbott Cell-Dyn 3000 Series Hematology Systems, Cell-Dyn Reticulocyte Reagent, when used in combination with Cell-Dyn 3500 reticulocyte software enables a whole blood specimen to be analyzed for reticulocytes. The Reticulocyte Reagent is intended for in vitro diagnostic use (outside the body). Recall #Z-715-9. REASON These lots may exhibit elevated Reticulocyte results. In some cases, the...
Read More
2/16/99 Class II 128XP Acuson 637 units. Rev 27 software with OB calculations options are affected. Potential error condition related to OB reports printed after diagnosis. Z-787-9
Read More
Solar 7000 and Solar 8000 Patient Monitors with Version 5D Software, used to display physiological data from modules which monitor the patient for ECG, blood pressure, cardiac output, respiration, pulseoximetry, etc. Recall #Z-727/728-9. REASON A software deviation can affect the alarm default settings, which may result in incorrect or missed alarm notification. CODE Solar 7000/8000...
Read More
10-Feb -99 Precision Therapy International. 190 units. Class II. 3-D Imgcomp Utility User Manual Update. Version 2.72b & 3.07d. A design error may leadto a serious miscalculation of radiation dose. Z-1223-9
Read More
2/9/99 Class II Bayer AADVIA120 Hematology System 920 units. The device may give erroneous reslts because of the valve gasket deterioration in the Baso channel check valve, and the software designed will not flag significant discrepancies in the results between the two methodologies at extremely low counts.Z-869/870-9
Read More
2/5/99 Class I recall Marquette Electronics ECG, blood pressure, cariac output, respiraion, pulse oximetry Monitors. 607 new units and 97 update kits. A software deviation can affect alarm default settings, which may result in incorrect or missed alarm notification. Z-727/728-9
Read More
2/1/99 Abbott Alcyon 300 Analyzer (IVD) Possible mixup of patient results. 631 units. Z-617-9
Read More
National Biomedical Computer System, Versions 1.1 to 1.3.2. Recall #B-861-9. REASON Y2K defect in the computer software incorrectly calculates the end date of a temporary deferral to be 1900 instead of 2000, and removes the deferral from the database. CODE Software versions 1.1 to 1.3.2. MANUFACTURER American Red Cross Blood Services, Arlington, Virginia. RECALLED BY...
Read More
Four Recalls of Picker International Nuclear Cameras and imaging systems12/15/98 Prism XPVT 63 units. Unexpected camera movement which requires the operator to power down and reboot with delay in diagnosis. Z-643/650-9 11/18/98 Model 3000 325 units. Unintended motion of gantry during setup. Z-641/642-9 11/17/98 ModelFX AND fx 800 WITH VERSION 8.3.11 software. 70 units. Software...
Read More
12/10/98 Datalink Management System Software Versions 4.1-4.2 Beckman Coulter 248 units. The devices may allow the test results from an associated clinical instrument to be placed under an incorrect sample ID. Z-829-9
Read More
PRODUCTVersion 6A and prior versions of software for Marquette Coherent Digital Telemetry (CDT) LAN Monitoring Systems, a patient monitoring system designed to collect and transmit ECG and other physiological data from ambulatory patients, without the patient being physically connected to a display device. Recall #Z-774-0.
Read More
11/24/98 HMT Medical Technologies Switzerland. Lithotripsy instruments Software versions 3.0 and 1.2. 17 units. Software upgrade yielding error messages and complete lock-up of instruments. Z-680/681
Read More
11/2/98 ACS Centaur Automated Chemiluminescence System 244 units with Software Version 1.2 The device may go into premature ready state with bleach still remaining in the analyzeer lines due to a software error. Z-499-9.
Read More
Software for the Abbott AxSYM System, a fully automated immunoanalyzer. Recall #Z-331-9. REASON Extra assay results from a previous run can appear with new sample results, as well as error messages, due to software defect in the internal result database management. CODE All software versions below 3.03. MANUFACTURER Abbott Laboratories, Irving, Texas. RECALLED BY Abbott...
Read More
9/4/98 Nera Electronics/Baxter Healthcare Flo Gard 8200 Volumetric Infusion Pump: Class II 13 units. Board assemblies had the wrong software version installed on them. The boards have version 2.09 instead of version 2.13. Recall #=-Z-777-8.
Read More
10/1/98 Class II Quinton Q-Cath Hemodynamic Analysis System 81 units. Units with version 4.5 software. Device has potential to display previous patient’s blood pressure. Z-855/859-9
Read More
GE Advantx Vascular Imaging Systems, angiographic X-ray systems: a) GE Advantx Vascular Imaging Systems (DX and DLX) with Stenosis Analysis Software and; b) Cardiac Review Stations (CRS) with Quantitative Analysis Software. Safety Alert #N-002/003-9. REASON When procedures in the Operator Manual for stenosis sizing are not followed each time a time measurement is taken, serious...
Read More
Blood Bank Control Systems. Recall #B-020-0. REASON Computer software contains programming errors that could potentially result in the release of a unit that has not been tested for antibodies. CODE Software versions 4.0 and greater. MANUFACTURER Blood Bank Computer Systems, Kent, Washington. RECALLED BY Manufacturer, by letter dated September 1, 1998. Firm-initiated field correction complete....
Read More
8/12/98 Class II PRODUCT Model No. 250 NeuroCybernetic Prosthesis Programming (NCP) Software, Version 5.0 for use only with the Model No. 100 NCP Pulse Generator and Model No. 200 NCP Programming Wand, an implanted autonomic nerve stimulator implanted for epilepsy. Recall #Z-492-0. REASON The software contains a flaw that reports an inaccurate amount of remaining...
Read More
7/20/98 Hewlett Packard Image Point Ultrasound System Model 2410A used in ultrasound imaging to determine stenosis of the sarotid artery. Software Revisions A.0.0 A.0.1, and A.0.1.2. Class II 441 units. Software overestimates peak blood velocity by 15-25% in some Doppler examinations with the Model 2135A transducer. Recall #=Z-160-8.
Read More
8/10/98 ACS Centaur Automated Chemiluminescence System 29 units with Software Version 1.0-1.2 The test results will be innacurate if the user changes the reporting units and/or slope and intercept for a test. Z-490-9.
Read More
6/25/98 Medtronic 9790C Programmers for Kappa 400 Pacemakekers Model 9952A software: Class II 133 units. Bench testing identified a software anomaly that could result in the delivery of an unitended pacing mode or pacing interval. Recall #=Z-715-8
Read More
1 59 60 61 62

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: March 7 & 28, 2024

Virtual via Zoom

Registration Link:

Register Now

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.