By

Amy Sellers
Company: Semens Medical SystemsDate of Enforcement Report: 7/31/02 Class: II PRODUCT SONOLINE Antares, Diagnostic Ultrasound System. The SONOLINE Antares is a high-resolution, digital, broadband diagnostic ultrasound imaging system. It supports the following diagnostic applications: abdominal, OB/GYN, small parts, musculoskeletal, vascular, and urology. Recall # Z-1162-2. REASON Software anomalies that affect measurement and calculation data. CODE...
Read More
Company: Fresenius Medical CareDate: 7/25/02 Product: Naturalyte Acid Concentrate Your firm failed to validate the new salt delivery system for process deviation 2002-005 dated February 1, 2002 for Naturalyte Acid Concentrate lot 2B0017 as required by 21 CFR 820.75. Furthermore, the new computer software included with the new salt delivery system is not validated for...
Read More
Company: Abbott LaboratoriesDate: 7/19/02 Product: Drug products including Biaxin and Meridia FDA compared the MedWatch forms in your computer to your paper files, and this comparison revealed inaccurate data in serious and unexpected adverse drug experience reports submitted to FDA. These inaccurate reports are a violation of 21 CFR 314.80(c)(1)(i). FDA District: Chicago District Office
Read More
Company: Varian Medical Systems, Inc Date of Enforcement Report: 7/3/2002 Class: II PRODUCT Varian Medical System’s Eclipse VERSION 6.5 software, a proprietary application for Varian Vision ™ imaging systems. Recall # Z-1109-2. REASON Radiation treatment planning software malfunction could result in incorrect treatment regimen. CODE Eclipse software, version 6.5, Serial Numbers: 0016 0018 0022 0034...
Read More
Company: Philips Medical SystemsDate of Enforcement Report: 7/3/2002 Class: II PRODUCT Information Center System and Upgrades as follows: a) M3150A Agilent or Virida Information center. Recall # Z-1119-2; b) M3151A Agilent or Virida Information Center Client. Recall # Z-1120-2; c) M3153A Agilent or Virida Surveillance Center. Recall # Z-1121-2; d) M3150AU#A01 Agilent or Virida Information...
Read More
Company: Toshiba America Medical Systems, IncDate of Enforcement Report: 6/26/02 Class: II PRODUCT NEMIO, SSA550A, Diagnostic Ultrasound System. Recall # Z-0923-2. REASON Software anomally caused error in calculating VM_P. CODE Serial Numbers: B1552349 C1572413 C1572414 C1572415 C1572417 C1572418 C1572420 C1572423 C1572428 C1572429 C1572430 C1572431 C1572432 C1572433 C1572421. RECALLING FIRM/MANUFACTURER Toshiba America Medical Systems, Inc., Tustin,...
Read More
483Company: American Red Cross City and Sate: Baltimore, MD FDA District: Baltimore, MD Dates of Inspection: 4/22-6/21/02 4. Failure of the quality control unit to determine root cause and prevent recurrence of significant deviations related to BSD 73.200M, “Shipping” version 1.7 January 2001. Between October 22, 2001 and April 23, 2002, the Chesapeake Region recorded...
Read More
Company: Advanced Sterilization ProductsDate: 6/19/2002 Class: II PRODUCT STERRAD 100S Sterilization System. Recall # Z-1082-2. REASON Software anomaly allows half cycle run when full cycle displayed. CODE Version 3 or lower is subject to upgrade. Newer versions contain software “fix”. Current version is Version 6. RECALLING FIRM/MANUFACTURER Advanced Sterilization Products, Irvine, CA, by letters sent...
Read More
Company: Fujifilm Medical Systems U.S.ADate of Enforcement Report: 5/29/02 Class: II PRODUCT Synapse Image and Information Management System. Recall # Z-1045-2. REASON Rotated/flipped Images not viewed on the screen may not rotate but left / right markers will be flipped. CODE Synapse Image and Information Management System with software versions 2.0.2 and 2.1.1. RECALLING FIRM/MANUFACTURER...
Read More
Company: Liebel-Flarsheim CoDate of Enforcement Report: 5/29/2002 Class: II PRODUCT Liebel-Flarsheim Angiomat Illumena Media Power Injectors, P/N 900001, using Version 8.0 software with new console. Recall # Z-1046-2. REASON Pushing the scoll button repeatedly / rapidly results in erroneous protocol ID. CODE Serial Numbers: 0202-5000 to 0402-5126. RECALLING FIRM/MANUFACTURER Recalling Firm: Mallinckrodt, Inc. / Liebel-Flarsheim...
Read More
Company: Sigma International MedinaDate of Enforcement Report: 5/8/02 Class: II PRODUCT Sigma 8000 Volumetric Infusion Pumps with software version 3.50.01. Recall # Z-0991-2. REASON Software error: Distal occlusions may not be detected (no alarm) at rates above 550 ml/hr. CODE The following serial numbers. 49105, 49111R, 49188R, 49288R, 49289R, 49402R, 49436R, 49448R, 49454R, 49466R, 49507R,...
Read More
Company: Hitachi LtdDate of Enforcement Report: 5/1/02 Class: II PRODUCT Roche Modular Analytics immunoassay based clinical chemistry analyzer, model E170, catalog number 7656961. Recall # Z-0922-2. REASON Possibility of false negative and false positive results being reported. CODE All units with software version 01-03. RECALLING FIRM/MANUFACTURER Recalling Firm: Roche Diagnostics Corp., Indianapolis, IN, by letter...
Read More
Company: Philips Medical SystemsDate of Enforcement Report: 5/1/2002 Class: III PRODUCT Integris H5000, Angiographic X-ray system. Recall # Z-0924-2. REASON System may occasionally re-boot during a patient examination. CODE Units with SYSCO software version 11.1.2. RECALLING FIRM/MANUFACTURER Recalling Firm: Philips Medical Systems North America, Inc., Bothell, WA, by letter dated November 14, 2001., Manufacturer: Philips...
Read More
Company: Roche Diagnostics CorpDate of Enforcement Report: 5/1/02 Class: II PRODUCT All DataCare Systems Software, as follows: Recall # Z-0960-2/Z-0974-2. a) DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9200; b) DataCare ABG Data Management System for use in a blood gas point of care testing...
Read More
Company: Sunquest Information systems, aka Misys HealthcareDate of Enforcement Report: 5/1/02 Class: II PRODUCT Misys Laboratory Microbiology Module, Recall # Z-0916-2. REASON Software does not allow comments to be shown in some reports. CODE Versions 5.2, 5.23 and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information systems, aka Misys Healthcare, Tucson, AZ. Firm initiated recall is ongoing. VOLUME...
Read More
Company: St. Gobain Ceramiques Avancees DesmarquestProduct: Medical Devices Date: 4/26/02 You need to property validate the … furnace process. Please supply a complete proposed validation protocol for the … furnace. This protocol should specifically address if any aspect of this furnace is computer controlled and how the software will be validated. Enclosed are copies of...
Read More
Company: Zoll Medical CorpDate of Enforcement Report: 4/24/02 Class: I PRODUCT a) Zoll M Series Advisory Defibrillator. Recall # Z-0893-2; b) Zoll M Series AED Defibrillator (semi automatic defibrillator). Recall # Z-0894-2. REASON Defibrillator may fail to detect ventricular fibrillation and fail to deliver shock. CODE a) Serial Numbers: T98F00046-T01K27762 with System Software Version below...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 4/24/02 Class: II PRODUCT Misys Laboratory for Windows version 5.3. Recall # Z-0888-2. CODE REASON Software QA code of BOTH for patient sex can result in QA flags not being generated for abnormal results. Versions 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., aka Misys Healthcare Systems Tucson,...
Read More
Company: American Dental Technologies Inc.Date: 4/18/02 Product: Dental device systems, including PulseMaster 600-IQ and DioLase ST Dental Laser System, PowerPAC High Speed Curing Light System, KCP 5 and 1000 Air Abrasive Kinetic Cavity Preparation System and Ultracam Intraoral Camera System Failure to maintain adequate device master records [21 CFR 820.181] [FDA- 483 Item 5]. For...
Read More
Company: A-VOX Systems Inc.Date: 4/18/2002 Product: Various oximeter models used for whole blood measurements and disposable cuvettes Failure to maintain procedures to ensure all purchased or otherwise received products and services conform to specified requirements [21 CFR 820.50; see also 820.80]. For example, your firm failed to ensure that the supplier of the main computer...
Read More
Class: III Aeroset Chloestorol Reagent Lot 82099HW00. Firm’s list no. 7D62-01. Manufacturer: Abbott Laboratories, Inc., Sout Pasadena, CA Recalled by: Manufacturer, by letters on April 17, 2002. Firm initiated recall is ongoing Distribution: Nationwide and internatinally; 715 units were distributed Reason: Stability may be compromised over time, product currently OK Recall number: Z-1031-2
Read More
Company: Misys Healthcare SystemsDate of Enforcement Report: 4/17/02 Class: II PRODUCT Misys Laboratory (formerly FlexiLab) information system. Recall # Z-0803-2. REASON Software anomaly may allow patient results to be transposed against another patient header or identification. CODE Versions 5.2, 5.23, and 5.3. RECALLING FIRM/MANUFACTURER Misys Healthcare Systems, Tuscon, AZ, by notice on March 8, 2002....
Read More
Company: Roche Diagnostics CorpDate of Enforcement Report: 4/10/02 Class: II PRODUCT All DataCare Systems Software, as follows: a) DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9200 b) DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9201...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 4/10/02 Class: II PRODUCT FlexiLab Laboratory Information System. Recall # Z-0710-2. REASON Date Format Issue–Flexilab Software converts specimen collect and receipt dates to American standard format ‘mm/dd/yyyy’. For sites using the Euro date format this can alter the generation of QA warnings or flags based on collect...
Read More
Company: Abbott Laboratories, Inc Date of Enforcement Report: 3/24/02 Class: II PRODUCT AEROSET System Software v1.00ER005.2 and v1.0ER005 Catalog Numbers: 2-8996-02 and 2-89996-01. Recall # Z-0594-2/0595-2. REASON System may incorrectly associate sample ID & test requests undercertain conditions of non-bar coded or unreadable bar coded samples. CODE All Codes are affected by this field correction....
Read More
Company: Bayer CorporationDate of Enforcement Report: 3/27/02 Class: III PRODUCT Clinitek 100 Software Kit. Recall # Z-0634-2. REASON False negative results when high concentrations of ketone exist. CODE Version 7.00. RECALLING FIRM/MANUFACTURER Recalling Firm: Bayer Corp., Elkhart, IN, by telephone and fax on January 21, 2002. Manufacturer: Bayer Corporation, Mishawaka, IN. Firm initiated recall is...
Read More
Company: Siemens AG Medical EngineeringDate of Enforcement Report: 3/27/02 Class: III PRODUCT a) ACOM.PC Angiographic X-ray System; b) Digital Image Communications System. Recall # Z-0603-2. REASON Software problem in recording heights and weights. CODE Siemens part number: a) ACOM.PC V3.0-4371105 b) Quantcor V4.0-4786898, 5915553, 5915579. Equipment Concerned: Product: Part No. Serial No. LVA V 4.0...
Read More
Company: Sunquest Information Systems IncDate of Enforcement Report: 3/6/02 Class: II PRODUCT FlexiLab Laboratory Information System, Version 5.2, 5.23, and 5.3, Recall # Z-0531-02. REASON Software glitch can cause results to be filed under an incorrect accession number under certain conditions. CODE Versions 5.2, 5.23, and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems Inc., Tucson, AZ,...
Read More
Company: Ochsner Clinical FoundationDate: 03/05/2002 Product: Investigational device The manual does not adequately describe how the CIC support staff prepares and maintains records of all IRB activities, including meeting minutes, records and reports associated with investigations, and archiving. The manual does not describe how data and information are gathered, stored, and analyzed to prevent research...
Read More
Company: ADAC LaboratoriesDate of Enforcement Report: 2/13/02 Class: II PRODUCT ADAC brand Pinnacle Radiation Therapy Planning System Radiation Therapy Treatment Software version 5.2g, Recall # Z-0524-02. REASON A Software problem that could result in incorrect radiation dosage being applied. CODE Model: 9200-0579A. RECALLING FIRM/MANUFACTURER ADAC Laboratories, Milpitas, CA, by letters on Sept. 26, 2000. Firm...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 2/13/02 Class: II PRODUCT FlexiLab? Laboratory Information System, Recall # B-0710-2. REASON Blood bank computer system, which had a software glitch or defect, was distributed. CODE Versions 5.23 and 5.3 with RBR module. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ, by letter dated March 19, 2001....
Read More
Company: Sunquest Information Systems, Inc Date of Enforcement Report: 2/13/02 Class: II PRODUCT FlexiLab? Blood Bank and Blood Donor Software, Recall # B-0711-2. REASON Blood bank computer system, which had a software glitch or defect, was distributed. CODE Versions 5.2 and 5.23. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ, by letter dated June 22,...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 2/13/2002 Class: II PRODUCT FlexiLab? Blood Bank and Blood Donor Software, Recall # B-0709-2. REASON Blood bank computer system, which had a software glitch or defect, was distributed. CODE Versions 5.2, 5.23 and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ, by letter dated March 13,...
Read More
Company: Roche Diagnostics CorpDate of Enforcement Report: 2/6/02 Class: II PRODUCT All DataCare Systems Software, as follows: A) DataCare Server/Workstation Computer for DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9200. Sold in the U.S. B) DataCare Networked Client Computer for DataCare ABG Data Management System...
Read More
Company: Dade Behring, IncDate of Enforcement Report: 1/30/02 Class: II PRODUCT Dimension Clinical Chemical Analyzer XL/RxL/ARx with software revision 5.2, Recall # Z-0433-02. REASON Instructions to software revision are missing a step. CODE Revision 5.2 The following serial numbers are involved in the recall: 971807, 973002, 970457, 973278, 970598, 970454, 970049, 971979, 973242, 970864, 973218,...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 2/30/02 Class: III PRODUCT Hemocare Blood Bank Data Management System Software, Recall # B-0626-2. REASON Computer software that could allow the assignment of an extended expiration date to blood products was distributed. CODE Versions 5.1, 5.2, 5.2a, and 5.2b. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY...
Read More
Company: Biochem Immunosystems (U.S.), IncDate of Enforcement Report: 1/23/02 Class: II PRODUCT Personal Lab Version 2.2a. The device is an Immunochemistry Analyzer Open System, Recall # Z-0463-02. REASON Samples were not picked up and added to the dilution tube. CODE Version 2.2a Catalog # PL01. RECALLING FIRM/MANUFACTURER Recalling Firm: Biochem Immunosystems (U.S.), Inc. Allentown, PA,...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 1/23/02 Class: II PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0670-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 1/23/02 Class: II PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0668-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 1/23/02 Class: 111 PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0671-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, Inc.Date of Enforcement Report: 1/23/02 Class: III PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0669-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, Inc.Date of Enforcement Report: 1/23/02 Class: III PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0667-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 1/23/02 Class: III PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0666-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 1/23/02 Class: II PRODUCT FlexiLab? Blood Bank and Blood Donor Software (3 versions), Recall # B-0627-2. REASON Blood establishment computer software that could allow the release of unsuitable blood products was distributed. CODE Version numbers 5.2, 5.23, and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ,...
Read More
Company: Sysmex CorpDate: 01/22/2002 Product name: Clinical laboratory diagnostic devices, including hematology, reticulocyte, urinalysis and coagulation analyzers Failure to establish and maintain documented procedures to control and verify the design of the device in order to ensure that specified design requirements are met, as specified in 21 CFR 820(a)(1). For example, there were no procedures...
Read More
Company: Siemens Medical Systems, IncDate of Enforcement Report: 1/16/02 Class: III PRODUCT Autoperfusion Software used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems to conduct automated myocardial processing in diagnosing myocardial disease in cardiac SPECT studies. Recall # Z-0427-02. REASON The software did not save oblique data correctly...
Read More
Company: Siemens Medical Systems, INCDate of Enforcement Report: 11/16/01 Class: III PRODUCT Autoperfusion Software, which includes The Cardiac Review (Profile Review) application, used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems to conduct automated myocardial processing in diagnosing myocardial disease in cardiac SPECT studies. Recall # Z-0428-02. REASON...
Read More
Company: Pharmaceutical Distribution SystemsDate: 01/03/2002 Product: Drug products Failure to exercise appropriate controls over and to routinely calibrate, inspect, or check automatic, mechanical, or electronic equipment used in the manufacturing, processing, and packaging of a drug product according to a written program designed to assure proper performance (21 CFR 211.68) in that, the installation qualification...
Read More
Company: Natural Technology Inc.Date: 12/14/01 Product: Drug products Failure to maintain records of calibration checks and inspections of automatic, mechanical, or electronic equipment [21 CFR 211.68(a)]. For example, incubator and refrigerator thermometers associated with equipment that stores microbiological media have not been calibrated. FDA District: Dallas District Office
Read More
Company: Medical Device Services Inc.Date: 12/13/2001 Product Name: Single use medical devices, such as general surgery, orthopedic, laparoscopic and cardiovascular instruments, including biopsy forceps, needles, burrs, drill bits, saw blades, trocars, retractors, blood pressure cuffs, catheters and guidewires Failure to validate computer software for its intended use according to an established protocol prior to approval...
Read More
1 49 50 51 52 53 58

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: March 7 & 28, 2024

Virtual via Zoom

Registration Link:

Register Now

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.