Company: Philips Medical Systems (Cleveland) IncDate of Enforcement Report:6/11/2014 Class lI: PRODUCT IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001; ISP IX, Model number 881030; ISP LX SPECT, Model number 881035), Philips Medical Systems. A medical software system for diagnostic imaging data viewing. Recall Number Z-1678-2014 REASON During SPECT reconstructions using Attenuation Correction and Scatter...
Read More
Company: Toshiba American Medical Systems IncDate of Enforcement Report:6/11/2014 Class lI: PRODUCT Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting. Recall Number Z-1714-2014 REASON Toshiba America Medical Systems, Inc. initiated a recall of Toshiba Infinix X-Ray (INFX-8000V/UU), because it has been...
Read More
Company: Physio-Control, Inc.Date of Enforcement Report:6/42014 Class lI: PRODUCT LIFEPAK 1000 defibrillator. The LP1000 (in AED mode) is indicated for use on patients in cardiac arrest. Recall Number Z-1660-2014 REASON Physio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to...
Read More
Company: NAGFA Corp.Date of Enforcement Report:6/4/2014 Class lI: PRODUCT IMPAX CV Pediatric Echo Reporting, IMPAX CV Echo Measurement Import via Optical Character Recognition (OCR) Service IMPAX. For the creation of digital structured reports for cardiovascular diagnostic studies such as echocardiography. Recall Number Z-1662-2014 REASON Failure to effectively install and test a hotfix to correct the...
Read More
Company Siemens Medical Solutions USA, Inc Date of Enforcement Report 6/4/2014 Class lI: PRODUCT ACUSON SC2000 Ultrasound Systems with software version 3.5 (VA35); ultrasound imaging system. Recall Number Z-1667-2014 REASON Software issues with ACUSON SC2000 ultrasound with software version 3/5 (VA35) – may incorrectly display the color flow data in the wrong position and when...
Read More
Company Siemens Medical Solutions USA, Inc Date of Enforcement Report 6/42014 Class lI: PRODUCT ACUSON SC2000 Ultrasound Systems with software version 3.5 (VA35); ultrasound imaging system. Recall Number Z-1668-2014 REASON Potential malfunction due to a software issue for ACUSON SC2000 ultrasound systems using software version 3.5 (VA35) when using the stress echo feature which could...
Read More
Company: GE Healthcare Date of Enforcement Report:6/4/2014 Class lI: PRODUCT The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professional, including radiologists, physicians, technologists, clinicians, and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. The...
Read More
Company Siemens Medical Solutions USA, Inc Date of Enforcement Report 5/28/2014 Class lI: PRODUCT syngo Imaging XS Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo Imaging XS also...
Read More
Company Siemens Medical Solutions USA, Inc Date of Enforcement Report 5/28/2014 Class lI: PRODUCT 550 TxT Treatment Table of the Digital Linear Accelerator (LINAC) To deliver X-ray radiation for therapeutic treatment of cancer. Recall Number Z-1645-2014 REASON Table may lose calibration during patient treatment, which may cause the possibility of mistreatment. RECALLING FIRM/MANUFACTURER Siemens Medical...
Read More
Company: Philips Healthcare Informatics, Inc.Date of Enforcement Report 5/28/2014 Class lI: PRODUCT Philips XRE, the dictation editor system included in the Philips XIRIS 8.1.50 XIRIS is a radiology information system (RIS) is a computer system that assist radiology services in the storing, manipulating and retrieving of patient information. This included patient scheduling and tracking and...
Read More
Company: GE Healthcare Date of Enforcement Report:5/28/2014 Class lI: PRODUCT GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01 Product Usage: The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location. The device is indicated for use by qualified medical personnel only. Recall Number Z-1640-2014 REASON When...
Read More
STERIS Corporation Product:facility sterilizes medical devices Date:5/22/2014 1. Failure to establish and maintain adequate procedures for implementing corrective and preventive action, as required by 21 CFR 820.100. Your CAPA procedure, PROC-00007, Revision 19, is deficient in that it does not adequately describe how to identify, correct and prevent the recurrence of nonconforming product and other...
Read More
Company: Baxter Corporation EnglewoodDate of Enforcement Report:4/2/2014 Class I: PRODUCT Baxter Corporation Englewood, ABACUS Total Parenteral Nutrition Calculation Software, Versions 3.1, 3.0, 2.1, and 2.0 – Software Errors May Cause Toxic or Overdose Symptoms. ABACUS TPN Calculation Software, Product Codes: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition),...
Read More
Company: SCC Soft Computer Date of Enforcement Report:5/21/2014 Class lI: PRODUCT SCC SA HIS version 4.0.7.0 when used with SoftMic version 4.0.7.0. SoftMic is a laboratory information system to be used in medical results, clinical, and public health laboratories by knowledgeable, trained, and experienced personnel. SoftMic enables the user to handle data processing and laboratory...
Read More
Company: Nellcor Puritan Bennett Inc. (dba Covidien LP)Date of Enforcement Report:5/21/2014 Class lI: PRODUCT Recalled product is a component of the Newport Medical Instruments e360 Ventilator: single board computer, part number SBC2105A. Intended to provide invasive or noninvasive ventilator support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency. Recall...
Read More
Company: Mckesson Medical ImagingDate of Enforcement Report:5/21/2014 Class lI: PRODUCT McKesson Radiology-PACS. McKesson Radiology is a medical image and information application that is intended to receive, transmit, store, retrieve, display, print and process digital medical images, digital medical video, and associated medical information from various medical imaging systems. Recall Number Z-1617-2014 REASON High priority alert...
Read More
Company: CareFusion 303, Inc.Date of Enforcement Report:4/23/2014 Class I: PRODUCT Alaris Pump Module (Model 8100), Software Version 9.1.18 The Alaris Pump Model 8100 is a large volume infusion pump. It is intended for use in health care facilities that use infusion for the delivery of fluids, medicines, blood, and blood products using continuous or periodic...
Read More
Company: Codman & Shurtleff, Inc Date of Enforcement Report:5/21/2014 Class lI: PRODUCT MedStream Programmable Infusion Pump, 20 mL, Product Code: 91-4200 The MedStream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intrathecal delivery of Morphine or Baclofen in the EMEA market. Recall Number Z-1571-2014 REASON A...
Read More
Company Siemens Medical Solutions USA, Inc Date of Enforcement Report 5/21/2014 Class lI: PRODUCT Syngo Dynamics Product Usage: Syngo Dynamics is a Picture Archiving and Communication System (PACS) intended for acceptance, transfer, display, storage, archive and manipulation of digital medical images, including quantification and report generation. Syngo Dynamics is not intended to be used for...
Read More
Company: INO Therapeutics Date of Enforcement Report:5/14/2014 Class lI: PRODUCT INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The INOmax DS delivery system delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The primary targeted clinical setting...
Read More
Company: Shimadzu Medical Systems Date of Enforcement Report:5/14/2014 Class lI: PRODUCT Shimadzu Digital Radiography X-Ray System, Catalog No. DAR-8000f. This device is intended to be used for radiography in the hospital with X-ray devices. Recall Number Z-1569-2014 REASON Shimadzu Corporation is recalling the Shimadzu Digital X-Ray System because they have received a report of a...
Read More
Company: GE Healthcare Date of Enforcement Report:5/14/2014 Class lI: PRODUCT Nellcor SpO2 Upgrade kit, consists of a Nellcor SpO2 circuit board, an OEM part manufactured by a third party supplier; and an integral component of the Nellcor SpO2 device. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with...
Read More
Company: Brainlab AG Date of Enforcement Report 5/7/2014 Class lI: PRODUCT ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X...
Read More
Company: GE Healthcare LLC Date of Enforcement Report:5/7/2014 Class lI: PRODUCT GE Healthcare, CardiacVX [CardiacVX (1.1.0 Ext. d) and CardiacVX Flow (1.1.0 Ext. d) analysis software running on Advantage Workstation]. Model #’s 5434389-2 and 5434384-2. The GE Cardiac VX for MRI is an analytical software tool, which provides reproducible tools for the review and reporting...
Read More
Sun Pharmaceutical Industries Limited Product: pharmaceutical manufacturing facility Date:5/7/2014 1. Failure to ensure that laboratory records included complete data derived from all tests necessary to ensure compliance with established specifications and standards. our firm frequently performs ?unofficial testing? of samples, disregards the results, and reports results from additional tests. For example, during stability testing, your...
Read More
Company: Hospira Inc..Date of Enforcement Report:5/7/2014 Class I: PRODUCT GemStar Docking Station, used with GemStar infusion pump, in combination with other GemStar accessories such as the GemStar Lock Box and patient bolus cord. The docking station is also used for hosting the GemStar pump for general infusions. List Numbers: 13075-XX-XX Recall Number Z-1512-2014 REASON There...
Read More
Company: Baxter Healthcare CorporationDate of Enforcement Report:2/7/2014 Class I: PRODUCT Sigma Spectrum Infusion Pumps with Master Drug Library Model No. 35700BAX and 35700ABB Affected products were manufactured from July 1, 2005 through January 15, 2014 and distributed from February 20, 2013 through January 15, 2014. Use: Sigma Spectrum Infusion Pumps with Master Drug Library are...
Read More
Company: RAYSEARCH LABORATORIES AB. Date of Enforcement Report 4/30/2014 Class lI: PRODUCT RaySearch Ray Station Radiation Therapy Treatment Planning System, Stand-alone software. Versions 2.0, 2.5, 3.0, 3.5 and 4.0. Recall Number Z-1481-2014 REASON RaySearch Laboratories has recalled “RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0” software due to dose miscalculations for bolus/external/fixation support structures...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 4/30/2014 Class lI: PRODUCT Philips MDC PACS R2.X / IntelliSpace PACS DCX R3.1 software, software application that is used for receiving, managing, archiving, distributing and recording medical images onto portable digital media (including but not limited to Compact Disk and DVD). The MDC PACS/IntelliSpace PACS DCX...
Read More
Company: Ortho-Clinical Diagnostics Date of Enforcement Report 4/30/2014 Class lI: PRODUCT VITROS Chemistry Products CDM PROM (Calibration Data Module) No. 0168, Catalog Number/REF 199 9077, used in conjunction with: VITROS Chemistry Products CRSC DT Slides, Generation (GEN) 82, REF 6802721, and VITROS DT Calibrator Kit, Product Code 1957924, and tested on the VITROS DT60/DT60 II...
Read More
Company:Ion Beam Applications S.A Date of Enforcement Report 4/30/2014 Class lI: PRODUCT Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116. Proton therapy Recall Number Z-1492-2014 REASON Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option...
Read More
Company: Boston Scientific Corporation Date of Enforcement Report 4/30/2014 Class lI: PRODUCT iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology. Recall Number Z-1487-2014 REASON Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 4/16/2014 Class lI: PRODUCT Philips HeartStart XL+ Defibrillator/Monitor with system software version B.00.00 or B.00.01 installed Model: 861290, automatic external defibrillator. Recall Number Z-1438-2014 REASON Software communication failure may occur on the HeartStart XL+ locking the user out of clinical mode and possibly causing a delay...
Read More
Company ZOLL Medical Corporation.Date of Enforcement Report:4/9/2014 Class lI: PRODUCT Zoll Medical X Series, PropaqMD, PropaqM Defibrillator/Pacemaker/Monitor With System Software Version 02.10.02.00 or Higher, automated external defibrillator. Recall Number Z-1311-2014 REASON Defibrillator/Pacemaker/Monitor May Latch in a Continuous Device Reset Loop and cannot be used RECALLING FIRM/MANUFACTURER ZOLL Medical Corporation, Chelmsford, MA on.3/52014. Voluntary: Firm Initiated...
Read More
Company RAYSEARCH LABORATORIES AB.Date of Enforcement Report:4/9/2014 Class lI: PRODUCT RaySearch Ray Station, Radiation Therapy Treatment Planning System, Stand-alone software.Recall Number Z-1308-2014 REASON RaySearch Laboratories has recalled RaySearch Ray Station 4.0 software due to clinical dose calculation errors during radiation therapy. RECALLING FIRM/MANUFACTURER RAYSEARCH LABORATORIES AB, Stockholm, Sweden on.2/6/2014. Voluntary: Firm Initiated recall is ongoing....
Read More
Company RAYSEARCH LABORATORIES AB.Date of Enforcement Report:4/9/2014 Class lI: PRODUCT RayStation Radiation Therapy Treatment Planning System, models 2.5, 3.0, 3.5 and 4.0; a software system designed for treatment planning and analysis of radiation therapy. Recall Number Z-1307-2014 REASON Potential for dose errors due to software program errors.. RECALLING FIRM/MANUFACTURER RAYSEARCH LABORATORIES AB, Stockholm, Sweden on.2/6/2014....
Read More
Company:Elekta, Inc.Date of Enforcement Report:4/92/2014 Class lI: PRODUCT Integrity 1.1. Integrity is the interface and control software for the Elekta range of medical digital linear accelerators and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes. Recall Number Z-1322-2014 REASON The problem is that the “static tolerances” from...
Read More
Company Spacelabs Healthcare, Llc.Date of Enforcement Report:4/9/2014 Class lII: PRODUCT TDWorkstation version 11.01.A or higher. TDWorkstation receives patient results from a connected instrument and sends them to a Laboratory Information System (LIS). Recall Number Z-1310-2014 REASON If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 4/9/2014 Class lI: PRODUCT Pinnacle3 Radiation Therapy Planning (RTP) System, software version 9.0. Recall Number Z-1312-2014 REASON Philips Medical Systems have recently determined that a software nonconformance can cause incorrect beam geometry. This issue affects Pinnacle3 Radiation Therapy Planning (RTP) System, software version 9.0 manufactured from...
Read More
Company Bio-Rad Laboratories, Inc.Date of Enforcement Report:4/2/2014 Class lI: PRODUCT D-10″ Hemoglobin Testing System, Model number: 220-0220 Manufactured by: Bio-Rad Laboratories, Inc. 4000 Alfred Nobel Drive Hercules, CA 94545 The D-10 Hemoglobin Testing System is a single module designed for hemoglobin analysis in mid-volume laboratories. The D-10 is a fully integrated system for sample dilution,...
Read More
Company Velocity Medical Solutions, LLC.Date of Enforcement Report:4/2/2014 Class lI: PRODUCT DICOM Export Format for Sum Dose with Velocity v3.0.0. A stand-alone software product that provides the oncology care specialists (physicians, physicists, dosimetrists, etc.) a means for comparison of medical imaging data from multiple DICOM conformant imaging modality sources. It allows the display, annotating, volume...
Read More
Company Spacelabs Healthcare, Llc.Date of Enforcement Report:4/2/2014 Class lI: PRODUCT Spacelabs Healthcare G2 Clinical Access Mislabeled CDs for PN: 063-1829-11 Rev A and Update Kit, PN: 040-1565-10. Used for installing ICS G2 Clinical Access software onto hospital personal computers. Recall Number Z-1273-201 REASON Spacelabs has voluntarily recalled G2 Clinical Access Software CD because the programs...
Read More
Company Medical Information Technology, Inc..Date of Enforcement Report:3/26/2014 Class lI: PRODUCT MEDITECH Microbiology; Client Server Releases 5.6, 5.65, 5.66, 5.67, 6.0, 6.05, 6.07, 6.1, 6.13 and 6.14. Integrates requisition, specimen processing, and patient information with data in the healthcare information system (HCIS). Recall Number Z-1234-2014 REASON Potential for erroneous result reporting.. RECALLING FIRM/MANUFACTURER Medical Information...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 3/262014 Class lI: PRODUCT Philips Model EPIQ 5 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of...
Read More
Company: GE Healthcare LLC Date of Enforcement Report:3/26/2014 Class lI: PRODUCT GE Healthcare Optima CT580/Discovery CT590RT scanners running software version 13BW16.3. The Discovery CT 590 RT and Optima CT 580 series of Computed Tomography systems are intended to produce cross-sectional images of the body by computer reconstruction of XRay transmission data taken at different angles...
Read More
Company: Siemens Healthcare Diagnostics, Inc.. Date of Enforcement Report:3/26/2014 Class lI: PRODUCT syngo Dynamics Picture Archiving and Communication System (PACS). Model numbers 10091804, 10091805, 10091806, 10091807, 10091673 Recall Number Z-1216-2014 REASON Potential patient data mixup. Improper error handling could allow for two different patients to be displayed in syngo Dynamics. Images shown for one patient...
Read More
Company:Elekta, Inc.Date of Enforcement Report:3/19/2014 Class lI: PRODUCT RetCam 3, RetCam Portable and RetCam Shuttle; manufactured by Clarity Medical Systems, Inc.; Pleasanton, CA 94588. General ophthalmic imaging including retinal, corneal and external imaging. RetCam 3, RetCam Portable and RetCam Shuttle; manufactured by Clarity Medical Systems, Inc.; Pleasanton, CA 94588. General ophthalmic imaging including retinal, corneal...
Read More
Company:Elekta, Inc.Date of Enforcement Report:3/19/2014 Class lI: PRODUCT XiO RPT System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on screen and in hard-copy, either two or three dimensional radiation dose...
Read More
Company: Philips Medical Systems, Inc. Date of Enforcement Report 3/19/2014 Class lI: PRODUCT Philips HeartStart MRx Monitor/Defibrillator Models: M3535A and M3536A with Q-CPR Meter Option B08 The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of...
Read More
Company: GE Healthcare LLC Date of Enforcement Report:3/19/2014 Class lI: PRODUCT The Innova 3131IQ (3131-IQ) and 2121 IQ (2121-IQ) biplane systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational angiography procedures. They are also indicated for generating fluoroscopic images of human anatomy...
Read More
1 21 22 23 24 25 59

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  June 24-26, 2025 (Virtual, Live)

For private, in-house courses, please contact us.

Email training@softwarecpr.com for more info.

 


 

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • TBD
See our post titled: 1st Quarter 2025 Agile Compliant Courses Scheduled

 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.