Company: Beckman Coulter Inc..
Date of Enforcement Report 12/23/2015
Class lI:
PRODUCT
UniCel DxH 600 Coulter Cellular Analysis System, Software Version 3.0.2.0, Part No. 775222 The UniCel DxH 800/DxH 600 analyzers are quantitative, multi-parameter, automated hematology analyzers for in vitro diagnostic use in screening patient populations found in clinical laboratories. The DxH 800 is available as an analyzer for use on a benchtop or with a floor stand. The DxH 600 is available as an analyzer for use on a benchtop only.
Recall Number Z-0433-2016
REASON
Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager..
RECALLING FIRM/MANUFACTURER
Beckman Coulter Inc., Brea, CA on 12/1/2015. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
834 units total (594 units in US)
DISTRIBUTION
Worldwide
___________________________________