Company:RAYSEARCH LABORATORIES AB.
Date of Enforcement Report:6/25/2014
Class lI:
PRODUCT
RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.
Recall Number Z-1806-2014
REASON
This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution..
RECALLING FIRM/MANUFACTURER
RAYSEARCH LABORATORIES AB, Stockholm, SE on 4/25/2014. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
25 sites/245 units (US), 25 sites/124 units (foreign)
DISTRIBUTION
Nationwide and Internationally