Company: RAYSEARCH LABORATORIES AB 
Date of Enforcement Report 10/12/2016
 Class lI:
PRODUCT
 
 RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0) — Radiation Therapy Treatment Planning System RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.
 Recall Number Z-0038-2017
REASON
 RaySearch became aware of the problem as it was discovered in cooperation with a customer experimenting with the dose calculation on a phantom. The correction concerns two issues found with the dose calculation when using a region of interest (ROI) of type Fixation or Support with material override within the patient outline (External ROI) in RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0)
RECALLING FIRM/MANUFACTURER
 RAYSEARCH LABORATORIES AB, Stockholm, Sweden on 10/4/2016 Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
 USA: 522 units, OUS 846 units
DISTRIBUTION
 Nationwide
___________________________________
