Cl Il Panorama Patient Monitoring Network

Company: Mindray DS USA, Inc., dba Datascope Patient Monitoring
Date of Enforcement Report 6/3/2009
Class:ll

PRODUCT
Panorama Patient Monitoring Network with ambulatory telepacks, and software version 8.5.5, 8.5.6, 8.6.1 or 8.7. Datascope Patient Monitoring Intended for use in a fixed location, in the healthcare facility setting, as a central viewing station. Recall # Z-1401-2009

REASON
Software Anomaly: Software anomaly can occur when the Panorama is attempting to play an audio file, either a physiological or technical alarm sound, at which time the system software may cause the system to freeze due to an interaction in the alarm system.

RECALLING FIRM/MANUFACTURER
Mindray DS USA, Inc., dba Datascope Patient Monitoring, Mahwah NJ, by letters, dated March 27, 2009. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
294 affected sites

DISTRIBUTION
Nationwide and Australia

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.