Cl II Lumenis Light Sheer Desire Diode Laser

Company: Lumenis Limited.
Date of Enforcement Report: 5/6/2015
Class lI:

PRODUCT

Light Sheer Desire Diode Laser System with XC Handpiece Accessory options.
Recall Number Z-1519-2015

REASON
Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended settings. Operator Manual parameters are lower than indicated for specific hair color and Fitzpatrick skin type resulting in insufficient treatment effect. May result in patient burns and hypopigmentation.

RECALLING FIRM/MANUFACTURER
Lumenis Limited, , IL on 3/25/2015 Voluntary: Firm Initiated recall is ongoing

VOLUME OF PRODUCT IN COMMERCE
117 units

DISTRIBUTION
Nationwide and Internationally

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software

ISO 14971 Risk Management (typically with heavy emphasis on software aspect)

IEC 81001-5-1 Cybersecurity Lifecycle

ISO 13485 QMS Requirements & optionally ISO 19011 for leading audits

Being Agile Yet Compliant (AAMI TIR 45)

IEC 29119-1 Software and System Testing

US Classification for Software Devices

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.