GemStar Infusion System, Class I Hospira Inc.

Company: Hospira Inc
Date of Enforcement Report: 11/25/2013
Class l:

PRODUCT

Hospira GemStar Infusion System
Models 13000, 13100, 13150, 13086, 13087, 13088
The GemStar Infusion System is a small, lightweight, single-channeled device designed for use in the home, hospital or anywhere electronic infusion is required. The device is intended for use in intravenous, arterial, subcutaneous, short-term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional foods and blood/blood products. The device can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, the GemStar Pump is used for ambulatory patients.

REASON
The proximal and distal pressure sensor calibration can drift resulting in the pump failing the Proximal or Distal Occlusion Operational Test, as described in the GemStar Technical Service Manual, or reporting one of the following errors during device setup or infusion:

Cassette Check ? D
Cassette Check – P
Proximal Occlusion
Distal Occlusion
Pressure Calibration Error
Bad Pressure Sensor Event
Bad Pressure Sensor State
Distal Pressure is Out of Range
Proximal Sensor is Out of Range
A pump with this issue may, instead of reporting an error, not detect occlusions or issue false occlusion alarms, which will stop the infusion and invoke visual and audible warnings to the user. This issue may also result in the pump shutting down.

RECALLING FIRM/MANUFACTURER
Hospira Inc., Lake Forest, IllinoiI on 3/15/2013. Voluntary: Firm Initiated recall is ongoing.

FDA District
Chicago District Office

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Upcoming Public Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software training course in Boise, Idaho. 3-Day course.

October 13-15, 2026

Limited Early Bird Registration: $995 / person
Non refundable but transferable.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.