Inc., , Cl IIl Cyberonics

Company: Cyberonics, Inc., ,
Date of Enforcement Report 2/6/2008
Class:lll

PRODUCT
Cyberonics, VNS Therapy System Model 250 Handheld programming system using Dell x5 handheld and preprogrammed with software v7.1 and higher, Recall # 0568-2008

REASON
Screen Freezes– The Dell X5 Handheld PC screen will freeze caused due to incompatibility between the Microsoft 2002 OS and the model Dell X5 handheld computer. Once frozen, the handheld device becomes non-responsive to user input.

RECALLING FIRM/MANUFACTURER
Cyberonics, Inc., Houston, TX, by letter on October 1, 2007 and November 15, 2007. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
6642 units

DISTRIBUTION
Nationwide and Internationally

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.