Company: Roche Diagnostics Operations, Inc.
Date of Enforcement Report: 11/13/2013
Class lI:
PRODUCT
COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling accessory A fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids.
Recall Number Z-0168-2014
REASON
Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the software by the due date. This software issue ONLY affects systems performing ISE tests.
RECALLING FIRM/MANUFACTURER
Roche Diagnostics Operations, Inc., Indianapolis, IN on 9/16/2013. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
527 units
DISTRIBUTION
Nationwide and Puerto Rico
___________________________________