Cl ll Beckman Coulter UniCel DxI 600 & 800

Company:Beckman Coulter Inc
Date of Enforcement Report 8/3/11
Class ll:

PRODUCT
1) Beckman Coulter UniCel DxI 800 Access Immunoassay System with UniCel DXI system software version 4.3. Part Numbers: A71456, 973100, A25288, A25285. Recall # Z-2685-2011;
2) Beckman Coulter UniCel DxI 600 Access Immunoassay System, Part Numbers: A71460 and A30260 with UniCel DXI system. Recall # Z-2686-2011

REASON
Test results may have incorrect ranges. If a default sample type setting on the tests screen is changed while running UniCel DXI system software version 4.3, the new setting is not saved to the system database. The changed sample type setting will reset to the previously saved setting when power is interrupted to the system, or if an operator reboots the system. If the default sample type is set to urine while running software version 4.3, the system will reset to the previously saved sample type setting, serum, when the power is interrupted or the system is rebooted. For an LIS test request with no specified sample type, the system applies the previously saved default sample type setting to the test request even if the system was not rebooted. A test result could be reported with ranges or a unit of measure corresponding to the previously saved sample type setting. Ranges may include references ranges, critical ranges, and LIS ranges defined on the DXI.

RECALLING FIRM/MANUFACTURER
Recalling Firm: Beckman Coulter Inc., Brea, CA, by letter on July 6, 2010.
Manufacturer: Beckman Coulter, Inc., Chaska, MN. Firm initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
794 units

DISTRIBUTION
Nationwide and Canada

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.