Company: Beckman Coulter Biomedical GmbH
Date of Enforcement Report: 7/31/2026
Class: II
PRODUCT
REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections.
Recall Number:
Z-2834-2026
REASON:
Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
RECALLING FIRM/MANUFACTURER:
Beckman Coulter Biomedical GmbH
VOLUME:
20 systems
DISTRIBUTION:
U.S.: FL, GA, IL, MI, and PA O.U.S.: Belgium, France, Hungary, Poland, Taiwan (Province of China), Tunisia, United Arab Emirates, and United Kingdom of Great Britain and Northern Ireland.