FDA Recall: Software issue may result in over- or under-delivery

Company: ICU Medical, Inc.
Date of Enforcement Report: 12/19/2025
Class: II

PRODUCT

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

Recall Number:

Z-0966-2026

REASON:

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

RECALLING FIRM/MANUFACTURER:

ICU Medical, Inc.

VOLUME:

12 customers

DISTRIBUTION:

US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

Upcoming Public Courses

No public courses are planned at this time.

If you would like to discuss a private, onsite course, please fill out the form below.

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.