FDA Recall: Software latency causing interrupted motion on display screen

Company: Virtual Incision Corporation
Date of Enforcement Report: 6/4/2025
Class: II

PRODUCT

MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3

Recall Number:

Z-1905-2025

REASON:

The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.

RECALLING FIRM/MANUFACTURER:

VIRTUAL INCISION CORPORATION

VOLUME:

3 units

DISTRIBUTION:

US Distribution to States: CA, FL, NE