FDA Recall: Autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.

Company: Medtronic Inc.
Date of Enforcement Report: 9/13/2024
Class II

PRODUCT

2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description: PROGRAMMER 209031 PORTABLE CRM; c) 2090AA, Product Description: PROGRAMMER 2090AA PORTABLE CRM; d) 2090AA4, Product Description: 2090AA4 PROGRAMMER-PORTABLE CRM; e) 2090AC, Product Description: SFTWR 2090AC PROGRAMMER-PORTABLE CRM W/; f) 2090AM, Product Description: PROGRAMMER CARELINK HONG KONG ASIA; g) 2090CHEN, Product Description: PRGM 2090CHEN PKG-ASSY PROGRAMMER 2090; h) 2090CHFR, Product Description: PRGM 2090CHFR PKG-ASSY PROGRAMMER 2090; i) 2090CL, Product Description: PROGRAMMER 2090CL INTL-JAPAN MKT; j) 2090EN, Product Description: PRGM 2090EN PKG-ASSY PROGRAMMER 2090 UK; k) 2090FR, Product Description: PRGM 2090FR PKG-ASSY PROGRAMMER 2090 IN; l) 2090LDEN, Product Description: PRGM 2090LDEN PROGRAMMER-PORTABLE LOANE; m) 2090NL, Product Description: PRGM 2090NL PKG-ASSY PROGRAMMER 2090 IN; n) 2090W, Product Description: PROGRAMMER 2090W TELEMETRY C; o) 2090WCHDE, Product Description: PROGRAMMER 2090WCHDE SWISS-GERMAN; p) 2090WCHEN, Product Description: PROGRAMMER 2090WCHEN SWISS-ENGLISH; q) 2090WCHFR, Product Description: PROGRAMMER 2090WCHFR SWISS-FRENCH; r) 2090WDE, Product Description: PROGRAMMER 2090WDE GERMAN MKT; s) 2090WFR, Product Description: PROGRAMMER 2090WFR FRENCH MKT; t) 2090WLD, Product Description: PROG 2090WLD ALL LANGUAGES W/ANALYZER; u) 2090WNL, Product Description: PROGRAMMER 2090WNL NETHERLANDS MKT

Recall Number:

Z-3152-2024

REASON:

Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.

RECALLING FIRM/MANUFACTURER:

Medtronic Inc.

VOLUME OF PRODUCT IN COMMERCE:

18788 units

DISTRIBUTION:

Worldwide distribution.

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses:

IEC 62304 and other Emerging Standards Impacting Medical Device Software

Being Agile & Yet Compliant

ISO 14971 SaMD Risk Management

Software Risk Management

Medical Device Cybersecurity

Software Verification

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.