FDA Recall: Software Missing Planning Stage When Definition is Outside of Reference Range

Company: Blue Belt Technologies, Inc
Date of Enforcement Report: 10/05/2023
Class II

PRODUCT

Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

Recall Number:

Z-0041-2024

REASON:

CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis

RECALLING FIRM/MANUFACTURER:

Blue Belt Technologies, Inc

VOLUME OF PRODUCT IN COMMERCE:

56 units

DISTRIBUTION:

US Nationwide distribution.