Recall- Software error resulting in cancelled nurse calls

Company: Baxter Healthcare Corporation
Date of Enforcement Report: 6/30/2022
Class II

PRODUCT

NaviCare Nurse Call/Voalte Nurse Ca

Recall Number:

Z-1306-2022

REASON

An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.

RECALLING FIRM/MANUFACTURER

Baxter Healthcare Corporation on 5/21/2022. Voluntary: Firm initiated

VOLUME OF PRODUCT IN COMMERCE

283 installations

DISTRIBUTION

Nationwide and Canada

About the author

Allison joined SoftwareCPR in 2022, bringing along over ten years of passion for building teamwork and growth in organizations. This is established in part by the greatest University the South has to offer- Mississippi State.

SoftwareCPR Training Courses

IEC 62304 and Emerging Standards Impacting Medical Device and HealthIT Software

ISO 14971 Risk Management (typically with heavy emphasis on software aspect)

IEC 81001-5-1 Cybersecurity Lifecycle

ISO 13485 QMS Requirements & optionally ISO 19011 for leading audits

Being Agile Yet Compliant (AAMI TIR 45)

IEC 29119-1 Software and System Testing

US Classification for Software Devices

IEC 62366 Usability Process and Documentation

Or just email training@softwarecpr.com for more info.

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.